background image

7

6

Chapter 7 : REPLACABLE PARTS

The replaceable parts and accessories of EMS devices are as given below 

Except leads, electrodes and battery, battery case cover, please do not try to re-

place the other parts of a device.

PARTS

01

ELECTRODES LEADS

02

ELECTRODES

03

9V BATTERY  ,TYPE 6F22

04

BELT CLIP

05

BATTERY CASE COVER

06

LEAD CONNECTOR

07

MAIN PCB

08

INTENSITY  KNOB

09

CONTRACTION  KNOB

10

RELAXATION KNOB

11

PULSE WIDTH KNOB

12

PULSE RATE KNOB

13

RAMP KNOB

Chapter 8 : ACCESSORIES

Each set AMS-4 EMS are completed with standard accessories and standard

label as given below

I. Accessories

REF. NO.

PRODUCT

Q

TY

AGF-101

40 X 40 MM Adhesive Electrodes

4 pieces

AGF-110

Electrodes Leads

2 pieces

9 V Battery

1 piece

Instruction Manual

1 piece

Carrying Case x 1 EA

1 piece

Chapter 6 : TECHNICAL SPECIFICATION

The technical specification details of  AMS-4 EMS are as follows.

MECHANISM

TECHNICAL DESCRIPTION

01. Channel

Dual, isolated between channels.

02. Intensity Control

Adjustable, 0-80 mA peak into 500 ohm load

each channel

03. Wave Form

Asymmetrical Bi-Phase Square Pulse

04. Voltage

0 to 34 V (Load :  500 ohm)

05. Power Supply

One 9V battery

06. Size

95(H) x 65(W) x 23.5(T) mm

07. Weight

115 grams with battery

 08. Pulse Rate

Adjustable, from 2 to 150 Hz

09. Pulse Width

Selectable 100uS, 200uS, and 300 uS

10. Contraction Time

Adjustable from 1-30 seconds

11. Relaxation Time

Adjustable from 1-45 seconds

12. Ramp Time

Adjustable from 1-8 seconds

13. Max. Charge per pulse 20 micro-coulombs

14. Battery Life

Approximately 25 hours at nominal settings.

PDF created with FinePrint pdfFactory trial version 

http://www.pdffactory.com

Содержание AMS-4

Страница 1: ...CMYK Port Washington NY 11050 V1 3 PDF created with FinePrint pdfFactory trial version http www pdffactory com...

Страница 2: ...es Lead Wires 10 12 Lead Wire Maintenance 11 13 Electrode Options 11 14 Electrode Placement 11 15 Tips For Skin Care 13 16 Application of Re usable Self Adhesive Electrodes 13 17 Adjusting the Control...

Страница 3: ...which would otherwise serve as a protective mechanism Chapter 1 INTRODUCTION EXPLANATION OF EMS Electrical Muscle Stimulation is an internationally accepted and proven way of treat ing muscular injur...

Страница 4: ...placed over the eyes in the mouth or internally 11 Keep this device out of the reach of children 12 Caution Federal law restricts this device to sale by or on the order of a physician Chapter 4 GENER...

Страница 5: ...1 piece Instruction Manual 1 piece Carrying Case x 1 EA 1 piece Chapter 6 TECHNICAL SPECIFICATION The technical specification details of AMS 4 EMS are as follows MECHANISM TECHNICAL DESCRIPTION 01 Cha...

Страница 6: ...c muscle group The EMS can be used in the circumstances to quickly break the spasm Use a quick pulse rate wide pulse duration and set the intensity to visible contraction still within patient toleranc...

Страница 7: ...patient can easily continue treatment at home CONTIGUOUS PLACEMENT This is the most common placement technique It involves placing the elec Stimulation will continue at the setting contraction time a...

Страница 8: ...firmly to the treatment site Removal 1 Lift at the edge of electrodes and peel do not pull on the lead wires because it may damage the electrodes trodes alongside the affected muscles or muscle group...

Страница 9: ...two lead connector The device must be switched off before connecting the cables Both intensity controls must be at the Off position Electrodes must be pressed firmly on the skin 2 Place the electrode...

Страница 10: ...witched to off position 5 Pulse Rate Control This dial determines how many electrical impulses are applied through the skin each second By turning these controls the number of current impulses per sec...

Страница 11: ...partment acid may leak from used batteries and damage the device Put the device and accessories in carrying box and keep it in cool dry place 6 The packed EMS device should be stored and transported u...

Страница 12: ...device for external damage deformation of the housing damaged or defective output sockets 2 Check the device for defective operating elements legibility of inscriptions and labels make sure the inscri...

Страница 13: ...PDF created with FinePrint pdfFactory trial version http www pdffactory com...

Отзывы: