- 17 -
except for surface cleaning and changing of test window
membrane. Don’t try to repair the instrument by yourself.
Drawell won’t be responsible for any damage caused by
your own repair and the warranty time prescribed on the
contract will be invalid at that time.
3.2 Change test window membrane
1
.
Disassemble the panel in front of the test window.
2
.
Remove old membrane.
3
.
Clean the back part of the panel and install the new membrane.
4
.
Adjust the position of the screw.
5
.
Fix the screw into the hole but not too tight.
3.3 Storage & Transportation
3.3.1 Storage
Regulations in nearly all countries will require that you store your analyzer in a
secured place to prevent access, use, and/or removal by unauthorized individuals.
Storage requirements will vary from country to country, particularly with regard to
storage at temporary job sites or away from your primary storage location, such as
hotels, motels and in vehicles. You should contact your local Radiation Control
Authority to identify the specific storage requirements in your jurisdiction.
3.3.2 Transportation
There are no radiation regulations released by US Department of Transportation
(DOT) or International Air Transport Association (IATA) regarding shipping of the
Beethor X3G Series analyzer. It is recommended that you ship the Beethor X3G
Series in its carrying case and add an over-pack to protect the sophisticated parts
inside the analyzer.
Do NOT ship the analyzer with the battery pack connected to the analyzer!