This alarm will trip if the suction superheat falls
below the value of Setpoint #67 for an allotted
time. When suction superheat drops too low,
liquid refrigerant could enter the compressor,
causing damage. If the compressor has low
suction superheat, a LOW SUCT SPHT alarm
will display. The unit will attempt to run again
after a few minutes, but another trip within the
next hour will result in a lockout.
Note: Supervisor password (2112) is
required to reset lockouts
MicroMag Alarms
Possible Causes
Solutions
Sensor fault on suction
pressure or temperature
See Appendix B to troubleshoot
Lack of a load
Verify the chiller has a heat load
Lack of fluid flow
Confirm active flow to the chiller
Ensure y-strainer (shown to the right) and
process plumbing are free of blockages or
restrictions, and that all ball valves are
opened
TXV
Verify superheat reading with physical gauges
Adjust the TXV (shown to the right)
Replace if needed
Refrigerant Overcharge
Contact Drake
Low Suction
Superheat