Have the instrument adjusted by the service personnel.
Calibration
Calibration may not be possible due to instrument and channel errors.
Carrying out the fresh air calibration
Calibrate the instrument to fresh air, free of measured gases or other interfering
gases. During the fresh air adjustment, the zero point of all sensors (with the
exception of the XXS O
2
and the CO
2
channel of the infrared sensor) is set to 0.
In the case of the XXS O
2
, the display is set to 20.9 vol. %.
Switch on instrument.
Press the
[+]
key three times to display the icon for fresh air calibration
» «
.
Press the
[OK]
key to start the fresh air calibration.
–
The measured values flash.
When the measured values have stabilized:
NOTICE
To check the measured value response times, apply t90 test gas to the X-am
via the calibration cradle. Check the results in accordance with the information
in the table in the enclosed supplementary documentation (order no.
90 33 890) until 90 % of the end display is reached.
NOTICE
After the bump test (menu), the display shows a printer icon even if there is no
printer connected to the bump test
station.
NOTICE
A potentially activated H
2
offsetting is automatically temporarily deactivated
during a manual calibration, a PC calibration or an automatic Bump Test for the
relevant duration.
NOTICE
Dräger recommends using the extended bump test for cross calibrations
(Dräger X-dock technical manual).
NOTICE
The fresh-gas adjustment/zero-point adjustment is not supported by the CO
2
channel of the infrared sensor and the XXS O
3
. A zero-point calibration of these
sensors can be made with the Dräger CC Vision PC software. A suitable zero
gas free from carbon dioxide and ozone (e.g. N
2
) must be used.
Press the
[OK]
key to perform the fresh air calibration.
–
The display containing the current gas concentration changes with the
display
»OK«
.
Press the
[OK]
key to exit the fresh air calibration function or wait for
approx. 5 seconds.
If a fault has occurred during the fresh air calibration:
–
The fault message
»
«
appears and »
« is displayed for the
respective sensor instead of the measured value.
In this case, repeat the fresh air calibration. If necessary, have the sensor
replaced by qualified personnel.
Adjusting the sensitivity for an individual
measuring channel
–
The span calibration can be carried out selectively for individual sensors.
–
In the case of the span calibration, the sensitivity of the selected sensor is
set to the value of the test gas used.
–
Use a standard test gas.
–
Allowed test gas concentration:
Connect the test gas cylinder with the calibration cradle.
NOTICE
A potentially activated H
2
offsetting is automatically temporarily deactivated
during a manual adjustment, a PC calibration or an automatic Bump Test for
the relevant duration.
Ex channel of the
infrared sensor
20 to 100 %LEL
1)
2)
/ 5 to 100 vol. %
1) 2)
1)
Depending on the selected data set.
2)
Depending on the measuring range and the measuring accuracy.
CO
2
channel of the
infrared sensor
0.05 to 5 vol. %
2)
O
2
10 to 25 vol. %
CO
20 to 999 ppm
H
2
S
5 to 99 ppm
H
2
HC
0.5 to 4.0 vol. %
NO
2
5 to 99 ppm
Test gas concentrations of other gases: see Instructions for Use of the
respective DrägerSensors.
Содержание X-am 5600
Страница 117: ...117 127 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger www draeger com ifu Dr gerSensor 90 33 890 X am 5600 X am 5600...
Страница 121: ...121 OK 3 3 2 1 ch4 UEG CAL 123 bt 2 A1 A2 TWA 1 STEL 1 H2S CO OK OK 3 2 1 1 Bump Test 5 21 O2...
Страница 122: ...122 A1 Dr ger IR Dr ger X am 5600 A1 O2 A1 A1 1 2 OK A2 OK STEL A2 O2 A1 A2 0 100 CH4 0 100 A2 STEL TWA...
Страница 146: ...146 EC LED Dr ger Dr ger X am 5600 A1 A1 A1 A1 LED A2 LED OK A2 OK A1 O2 A2 O2 A1 A2 0 100 Vol CH4 0 100 LEL...
Страница 151: ...151 OK OK OK 5 EN 60079 29 2 EN 45544 4 4 Ex O2 H2S SO2 NO2 CO 6 Dr gerSensor O2 Dr ger 83 18 639 Dr ger 3 Dr ger Dr ger...
Страница 179: ...179 OK OK 3 2 1 5 m 21 Vol 1 A1 Dr ger IR A1 O2...
Страница 180: ...180 A1 A1 A1 A2 OK A2 OK STEL TWA STEL TWA A2 O2 A1 A2 CH4 0 100 Vol 0 100 LEL A2 STEL TWA STEL 1...
Страница 181: ...181 OK 20 10 Dr ger OK 3 OK TWA STEL 10 2 1 3 PC Dr ger CC Vision 1 OK 60 1 CO2 XXS O3 PC Dr ger CC Vision N2...
Страница 183: ...183 0 5 l 83 18 752 IR Ex 20 1 IR CO2 20 1 O2 0 6 Vol 1 TOX 20 1 A1 A2 H2 PC H2 1 Dr ger 68 11 132 t90 90 33 890 90...
Страница 187: ......