For applications in accordance with CSA (Canadian Standards
Association), the following should be observed:
Only the performance of the detector part of this instrument for flammable gases
has been tested. The instrument has not been approved for use in mines by the
CSA.
WARNING
The sensitivity must be tested on a daily basis before first use with a known
concentration of the gas to be measured in accordance with 25 to 50 % of the
concentration limit value. The accuracy must be 0 to +20 % of the actual value.
The accuracy can be corrected via calibration.
What is what
Special symbols:
1
Gas entry
8
IR interface
2
Alarm LED
9
Fastening clip
3
Buzzer
10
Type plate
4
[OK]
key
11
Charging contacts
5
Power pack
12
Measured gas display
6
[ + ]
key
13
Measured value display
7
Display
14
Special symbols
15
Tool for replacing the sensor
Fault message
1-button calibration
Warning
Single gas calibration
Display peak value
Password required
Show TWA
Battery 100 % full
Show STEL
Battery 2/3 full
Bump test mode
Battery 1/3 full
Fresh air calibration
Battery empty
00123896.eps
2
1
7
6
4
5
3
2
15
9
11
12
13
14
0
10
8
ch
4
%LEL
O
2
Vol%
CO
ppm
H
2
S
ppm
NH
3
ppm
X-am 5000