Section 3
Routine Maintenance
NIBP
Maintenance
Screen
To enter the NIBP Maintenance Screen:
1. Use the rotary dial to position the cursor on the
NIBP
selection.
2. Press the rotary dial.
This screen indicates the cuff inflation pressure and also provides a timing
period after inflation for leak testing. On-screen instructions provide a
method of checking pressure calibration with an external gauge. Refer to
the figure below.
1. Connect a BP cuff to the interface panel as instructed on the screen.
2. Wrap the cuff loosely around a cylindrical object.
3. Use the rotary dial to position the cursor to
TEST.
4. Press the dial to begin cuff inflation.
5. Wait one minute, then compare gauge pressure to
CUFF PRESSURE
VALUE
. The readings should agree within ± 5 mm Hg.
The Cuff Pressure Value will be displayed until the cuff deflates
(approximately four minutes after the start of inflation). The
STATUS
line
will then display the result of the test.
MAIN
PMS
SCHEDULE
MONITORS
KEYS
ROTATE DIAL TO MOVE CURSOR
PRESS TO ACTIVATE FEATURE
SU30043
TEST
SPO2
NIBP
SAMPLE
LINE
GAS
ANALYZER
-PLACE T FITTING INLINE WITH THE BP CUFF
-CONNECT A CALIBRATED PRESSURE GAUGE TO T
-SELECT 'TEST' KEY AND PRESS THE ROTARY
DIAL TO INFLATE CUFF
-VERIFY THAT THE CUFF INFLATES TO 300 MM HG
-WAIT FOR STATUS MESSAGE TO APPEAR BELOW
CUFF PRESSURE VALUE: 300
NIBP CUFF PRESSURE CHECK PROCEDURE:
STATUS: PASS
3-7
Содержание Vitalert 3000 Series
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 4: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 16: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 36: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 38: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 39: ...RETURN TO CD ROM TABLE OF CONTENTS ...