Section 3
Routine Maintenance
Following PMS month selection:
1. Use the rotary dial to position the cursor on
MONITORS
to proceed
with the other maintenance screens. (Selecting
MAIN
will return you to
the Main Maintenance Screen.)
2. Press the dial to return to the selected screen.
SpO
2
Maintenance Screen
To enter the SpO
2
Maintenance Screen:
1. Use the rotary dial to position the cursor on the
SPO2
selection.
2. Press the rotary dial. A typical screen is shown in the figure below.
This screen displays current values for SpO
2
and Pulse to verify that the
pulse oximeter is operational. A calibration check can be made by
disconnecting the SpO
2
sensor prior to power up, and connecting a
Nellcor® model PT-2500 Pocket Tester. The screen should then display an
SpO
2
value of 81 ± 1, and a Pulse value of 61 ± 1.
To return to the main maintenance screen or select another maintenance
screen:
1. Use the rotary dial to position the cursor on the
MAIN
or another
maintenance selection.
2. Press the dial to enter the selected screen.
MAIN
PMS
SCHEDULE
MONITORS
KEYS
ROTATE DIAL TO MOVE CURSOR
PRESS TO ACTIVATE FEATURE
SU30042
SPO2
NIBP
SAMPLE
LINE
GAS
ANALYZER
SPO2 VALUE: 81
PULSE VALUE: 61
3-6
Содержание Vitalert 3000 Series
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 4: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 6: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 16: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 36: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 38: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 39: ...RETURN TO CD ROM TABLE OF CONTENTS ...