12
Dräger VarioGard 2300 / 2320
Installation
3
Installation
Details on the mechanical installation and the wiring diagram are
available in the VarioGard 23x0 installation instructions (order no.
90 33 469).
4
Operation
VarioGard 23x0 transmitter
For monitoring of the ambient
air, depending on the sensor
installed.
Three-coloured optical LED.
Contact points for operation
using a magnetic wand.
5
Configuration
Use the PC software VarioCal to prepare calibration and to adjust gas
settings.
6
Maintenance
Visual inspection
At least every 3 months to determine operational readiness.
Routine inspections
Dräger recommends having the system inspected by Dräger Service
at least once a year.
When necessary
Replace the sensors of the transmitters.
Calibration of transmitters
The calibration can be carried out with the magnetic wand (order no.
45 43 428) directly on the transmitter.
Handling the magnetic wand:
Each transmitter has two contact
points labelled with
and
on the housing. Place the
magnetic wand on the contact
points for calibration.
00233185.eps
V
arioGard
CAUTION
Always calibrate the zero-point before span.
!
00833185.eps
V
arioGard
Содержание VarioGard 2300
Страница 2: ......
Страница 52: ...52 Dr ger VarioGard 2300 2320 1 Dr ger Dr ger Dr ger Dr ger 2 VarioGard 2300 2320 VarioGard 23x0 i i...
Страница 54: ...54 Dr ger VarioGard 2300 2320 6 1 VarioGard 23x0 USB 83 22 016 VarioCal 5 3 0 A 2 VarioGard 23x0 0 5...
Страница 55: ...Dr ger VarioGard 2300 2320 55 2 2 VarioGard 23x0 5...
Страница 56: ...56 Dr ger VarioGard 2300 2320 6 2 VarioGard 23x0 USB 83 22 016 VarioCal 5 2 VarioGard 23x0 0 5...
Страница 57: ...Dr ger VarioGard 2300 2320 57 2 1 2 VarioGard 23x0 5...
Страница 58: ...58 Dr ger VarioGard 2300 2320 6 3 6 4 VarioGard 23x0 2 VarioGard 23x0 2...
Страница 83: ......