15
Maintenance
Action
Status
display
Output
signal
Meaning
Carry out span adjustment:
Place the magnetic wand
on the marking »
« for
2 seconds and then
remove it again.
Green flashes Maintenance
mode
Being adjusted.
Stop the calibration gas.
Remove the calibration
adaptor VarioGard 23x0.
Ensure that the
transmitter is no longer
being purged with
calibration gas.
When the calibration has been completed, the
transmitter must be switched back to
measurement mode using the magnetic wand.
Place the magnetic wand on the marking »
«
for more than 5 seconds and then remove it again.
The calibrations that were successfully carried out
will be accepted.