Preparing and maintaining the 

saver

 CF

The tasks in this section must be carried out to prepare and maintain the 

saver

 CF in a ready-for-use condition.

4.1  Maintenance periods

4.1.1  Daily checks

It is essential that escape apparatus is ready for use at all times, and Dräger therefore recommend a daily check. If the customer’s 

on-site risk assessment concludes that less regular checks are acceptable, this can be extended to a maximum of one month. It is the 

customer’s responsibility to ensure that the equipment is ready for use at all times.

●  Check that the pointer of the cylinder pressure gauge is inside the green area.

●  Check the anti-tamper tag(s) on the lid of the bag (Fig 3).

If the pressure gauge is in the red area or if an anti-tamper tag is broken, inform trained service personnel or Dräger immediately.

4.1.2  Maintenance table

Dräger recommend that regular inspection, testing and servicing of the breathing apparatus is carried out in accordance with the table 

below. The table applies also to out-of-use (stored) equipment.

Additional inspection and testing may be required in the country of use to ensure compliance with national regulations governing the 

use, maintenance, examination and testing of this product.

4.3  Cleaning and disinfecting

 

CAUTION

 

Products  other  than  the  recommended  cleaning  and 

disinfecting  agent  are  not  tested  and  may  damage  the 

equipment. Contact Dräger before using any other cleaning 

or disinfecting agent.

 

When  using  cleaning  and  disinfecting  agents,  refer  to 

manufacturers’  instructions,  paying  particular  attention  to 

concentration and reaction times.

 

Do  not  exceed  30  °C  for  washing  and  rinsing  solutions. 

Do not exceed 60 °C or 30 minutes for drying, and remove 

components from the dryer immediately when dry.

 

Do  not  immerse  the  carrying  system  or  pneumatic 

components in cleaning solutions or water.

Recommended cleaning and disinfecting materials

:

●  Cleaning and disinfecting agent – Sekusept. 

●  Disinfecting agent – Incidur.

●  Use only or clean lint-free cloths.

1.  Clean  and  disinfect  the  carrying  system  and  pneumatic 

components manually using a cloth moistened with cleaning 

and disinfecting solution.

2.  Remove  all  cleaning  and  disinfecting  solution  using  a  cloth 

moistened with water.

3.  Dry all components using a dry cloth.

4.  Contact service personnel or Dräger if more thorough cleaning 

or disassembly of the pneumatic system is required.

Troubleshooting

There  is  no  user  troubleshooting  on  the 

saver

  CF.  Contact 

trained service personnel or Dräger to report any issues with the 

product.

Storage

Store the 

saver

 CF in a cool dry environment, free from dust and 

dirt, in a location where it is not subject to wear or damage due to 

abrasion. Note also the following:

●  Extend the neck strap and, if fitted, the waist belt. On the hard-

case carrying bag, the neck strap can be fed into the bag at the 

front and rear slots to tidy the strap.

●  Fix  the  apparatus  securely  to  any  raised  mounting  point  to 

prevent it from falling.

●  Contact Dräger for suitable storage cabinets and wall mounting 

kits.

Disposal

When  required,  dispose  of  the 

saver

  CF  in  accordance  with 

national or local regulations for waste disposal.

Order list

!

Component/System

Task

After use

Every 

year

Every 10 

years

Complete equipment

Visual inspection (see Section 4.2.1)

Clean and disinfect (see Section 4.3)

Functional tests (see Note 1)

Cylinder valve/reducer

Basic overhaul. Contact Dräger for the Repair Exchange Service (REX)

Cylinder

Charge to correct pressure (see Section 4.2.2)

Cylinder pressure test and recertification; check the test date on the cylinder

Carry out in line with national 

regulations

Notes

  Dräger recommendations 

1  These maintenance tasks may only be carried out by Dräger or trained service personnel. Details of the tasks are contained in the 

technical manual which is issued to service personnel that have attended a relevant Dräger maintenance course.

4.2  Maintenance tasks

4.2.1  Visual inspection

Carry out a visual inspection, checking the full breathing apparatus 

including  all  component  parts  and  accessories.  Check  that  the 

equipment is clean and undamaged, paying particular attention to 

pneumatic  components,  hoses  and  connectors.  Typical  signs  of 

damage  that  may  affect  the  operation  of  the  breathing  apparatus 

include impact, abrasion, cutting, corrosion and discoloration. Report 

dirty or damaged equipment to trained service personnel or Dräger, 

and do not use until faults are rectified and the product is cleaned.

4.2.2  Air cylinder charging

 

NOTICE

 

The charging procedure in this section describes recharging 

a cylinder that has been fully discharged during use of the 

saver

 CF. If during a check the cylinder pressure gauge is 

found to be in the red area, do not use this procedure and 

inform trained service personnel or Dräger immediately.

 

WARNING

 

Air  quality  for  compressed-air  cylinders  shall  meet  the 

requirements for breathing air according to EN12021.

●  The charging connector is a G5/8 connector as per EN 144-2.

●  Refer also to the instructions supplied with the cylinder and the 

charging unit for recharging the cylinder.

●  Only charge compressed-air cylinders which:

  ○  Conform to national standards.

  ○  Feature the original manufacturer’s test date and test mark.

  ○   Have not exceeded the test date indicated on the cylinder 

by the last testing station.

  ○  Are not damaged.

●  Dräger recommend a charge rate of 27 bar/minute (rapid charging 

will increase the temperature resulting in an incomplete charge).

●  To  prevent  overcharging  of  the  cylinder,  Dräger  recommend 

using a pressure-limiting device on the charging compressor.

1.  If the carrying bag is closed, open the lid and remove the hood 

from the bag.

2.  Remove the air cylinder as follows:

  ○   Soft-case  carrying  bag:  release  the  Velcro  strap  and 

carefully remove the cylinder.

  ○   Hard-case  carrying  bag:  remove  the  screws  from  the 

carrying bag body (Fig 5) (5 mm hexagon key and a 10 mm 

socket required), fully open the carrying bag body and then 

carefully remove the cylinder. Remove the lid from the slot 

on the bag.

i

i

!

3.  Using  the  probe  (Fig  6,  Item  1)  (Dräger  part  no.  3310488), 

press  the  plunger  down  and  insert  the  locking  clip  (Fig  6, 

Item 2).

4.  Remove the protection cap from the charging port and connect 

the charging adaptor (Fig 7).

5.  Recharge  to  the  rated  working  pressure  of  the  cylinder 

(200 bar).

 

NOTICE

 

Compression of cylinder air can cause a small temperature 

increase during refilling, resulting in an incomplete charge. 

If necessary, top-up the cylinder when it cools.

 

As charging commences, the whistle will sound briefly until 

the cylinder pressure reaches approximately 10 bar.

6.  When  the  cylinder  is  fully  charged,  vent  pressure  from  the 

charging hose and then remove the charging adaptor from the 

valve.

7.  Refit the air cylinder as follows:

  ○  Soft-case carrying bag:

    i.  Fully insert the cylinder into the carrying bag.

    ii.   Fit the pressure gauge shroud into the cut-out section 

in the side of the bag.

    iii.   Connect the Velcro strap around the cylinder.

  ○  Hard-case carrying bag:

    i.   Place the cylinder on the fully open case (Fig 8) and fit 

the neck strap as shown.

    ii.   Close the carrying bag body, ensuring that the pressure 

gauge shroud remains in its holder.

    iii.   Fit  the  securing  screws  (5  mm  hexagon  key  and  a 

10 mm socket required). Dräger recommend a torque 

of 2 Nm (1.5 lbf ft). The nylon in the locknut must exhibit 

an interference with screw threads – if interference is 

not felt, replace with a new M6 self-locking nut.

    iv.  Refit the lid into the slot on the bag.

 

NOTICE

 

Route the rubber hose in such a way that the bend radius 

is not too acute and the hose is not stretched, compressed 

or twisted.

8.  Carefully fold the escape hood to ensure that the visor is not 

creased, and fit the hood into the carrying bag (Fig 9).

9.  Seal the lid along the Velcro fastening strip leaving sufficient 

opening to insert a hand and, taking care not to detach the 

locking clip from the cylinder valve, connect the sprung-gate 

clip (Fig 1, Item 2) to the D-ring (Fig 1, Item 1).

10. Fit the anti-tamper tag(s) and close the lid (Fig 3).

i

i

i

i

Description

Quantity

Order code

Sekusept cleaner

4 x 2 litres

7904071

Incidur disinfectant

6 litres

7904072

Incidur disinfectant

30 litres

7904073

Anti-tamper tags

5

3350388

Waist belt

1

3350396

Self-locking nut (M6 with nylon insert)

50

1333062