Dräger Novametrix Скачать руководство пользователя страница 9

INSTALLATION PROCEDURE (continued)

Novametrix Monitor SpO2 Calibration Check:

1.

Plug the SpO2 sensor extension cable into the machine interface.

2.

Disconnect the finger probe from the extension cable and attach a Novametrix
TB500B sensor simulator to the extension cable.

3.

Turn the System Power switch to ON. press and hold the Selection Dial, and
simultaneously press the MONITOR and SYSTEM CONFIG keys to enter the main
service screen. Touch the MONITORS key to bring up a second column of touch keys.

4.

Touch the SpO2 key on the screen.

5.

Set the simulator’s Sensor Type switch to 87XX. Set the output signal to 3 and the
saturation to 82%. Turn the simulator ON and verify that the monitor’s displayed
Pulse reading is 60 ±1 BPM and the SaO2 reading is 82 ±2.

NOTE:

Do not rely on the simulator’s low battery LED as an indicator of sufficient
battery power. Replace the battery with a fresh 9 volt transistor battery if
proper measured values are not obtained and the battery measures

8.5 VDC.

6.

Adjust the simulator’s saturation setting to 100 and verify that the monitor indicates
99 ±1. Adjust the saturation to 62 and verify that the monitor indicates 62 ±2.

7.

Depress the Open Test RED key and verify that the PULSE LO LED activates, a
continuous audible alarm is heard, and a NO OXI PULSE warning alarm is shown
on the central alarm display.

8.

Depress both Open Test keys together and verify that the monitors SaO2 and OXI
ALARMS DISABLE indicators activate, a single audible tone is heard, and OXI SEN
DISC advisory alarm is shown on the central alarm display. Release the Open Test
keys.

9.

Disconnect the simulator and re-connect the finger probe. Attach the sensor to the
operator’s finger and obtain pulse, and oxygen saturation readings. Disconnect the
sensor from the finger. Afther ten (10) seconds, the Warning message NO OXI PULSE
shall appear on the central alarm display and a continuous audible alarm shall sound.

10.

Re-connect the sensor to the finger.

11.

Note the pulse high and low alarm settings. After pulse and oxygen saturation
readings stabilize, raise the pulse low alarm limit above the actual pulse rate.

12.

The Warning message OXI PULSE LOW shall appear on the central alarm display
and a continuous audible alarm shall sound.

13.

Return the pulse low alarm to its previous settings.

14.

Decrease the pulse high alarm limit below the actual pulse rate.

8

RETURN TO CD-ROM TABLE OF CONTENTS

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS

Содержание Novametrix

Страница 1: ...ld Service Procedure Part Number SP00173 Rev B Date 30 December 1999 1999 Draeger Medical Inc Installation of Novametrix SpO2 Module in Narkomed 4 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 2: ...ng items from the patient interface panel on the left side of the monitor box Remote Display Datagrip cables Pulse Oximeter Sensor cable BP Cuff and BP Gauge lines 3 Remove the disposable reservoir from its holder 4 Remove the screws holding the monitor box cover and carefully lift the cover from the monitor box 5 Disconnect all backplane wire harnesses and cables to the SpO2 assembly and disconne...

Страница 3: ...OX TOP VIEW WITH COVER REMOVED E T SV40438 PULSE OXIMETER ASSEMBLY PULSE OXIMETER ASSEMBLY SPO2 SPEAKER AND AMPLIFIER LATER DESIGN Figure 1 PULSE OXIMETER ASSEMBLY LOCATION 2 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 4: ...BACKPLANE J43 SV40437 EARLIER DESIGNS LATER DESIGN CONNECT WIRE HARNESS TO BACKPLANE J21 AND BACKPLANE J43 CONNECT RIBBON CABLE TO BACKPLANE J12 CONNECT CABLE FROM SpO2 INTERFACE PANEL J1 JP5 JP3 CONNECT CABLE FROM SpO2 INTERFACE PANEL CONNECT CABLE TO BACKPLANE J12 JP5 JP3 Figure 2 NELLCOR PULSE OXIMETER ASSEMBLIES 3 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS...

Страница 5: ...INSTALLATION PROCEDURE continued Figure 3 NOVAMETRIX PULSE OXIMETER MODULE 4 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 6: ...ard the bottom of the chassis captive screws 7 3B Proceed to Step 8 The following steps apply to the later design SpO2 assembly See Figure 1 and Figure 2 7 1C Remove the four mounting screws securing the SpO2 module to the chassis and remove the module 7 2C Install the Novametrix SpO2 module to the bracket with the hardware supplied in the kit Position the new module in the same manner as the orig...

Страница 7: ...isplay cable BP cuff and BP gauge lines and the oximeter sensor cable at the left side panel of the monitor box Replace the disposable reservoir 15 Add the Operator s Instruction Manual Supplement P N 4114884 to the Operator s Instruction binder 16 Restore power to the machine 17 Perform a PMS procedure on the machine including the Novametrix Monitor calibration check given on the following pages ...

Страница 8: ... simulated Use 86XX for non SuperBrightTM monitors and 87XX for SuperBright 2 RED open push button to simulate red LED failure 3 INFRARED open push button simulates open infrared LED 4 EXTERNAL POWER Jack is used to connect power from an external 9 VDC source When powered externally the simulator s internal battery is disconnected 5 SATURATION selects the simulated saturation value 6 POWER On Off ...

Страница 9: ...y that the monitor indicates 62 2 7 Depress the Open Test RED key and verify that the PULSE LO LED activates a continuous audible alarm is heard and a NO OXI PULSE warning alarm is shown on the central alarm display 8 Depress both Open Test keys together and verify that the monitors SaO2 and OXI ALARMS DISABLE indicators activate a single audible tone is heard and OXI SEN DISC advisory alarm is sh...

Страница 10: ...ssage SpO2 SAT LOW shall appear on the central alarm display and a continuous audible alarm shall sound 19 Return the SAT low alarm to its previous setting 20 Decrease the SAT high alarm limit below the actual SpO2 level 21 The caution message SpO2 SAT HIGH shall appear on the central alarm display and an intermittent audible alarm shall sound 22 Return the SAT high alarm to its previous setting 2...

Страница 11: ...North American Dräger DrägerServiceSM 3122 Commerce Drive Telford PA 18969 215 721 5402 800 543 5047 215 723 5935 Fax RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Отзывы: