INSTALLATION PROCEDURE (continued)
15.
The caution message OXI PULSE HIGH shall appear on the central alarm display
and an intermittent audible alarm shall sound.
16.
Return the pulse high alarm limit to its previous setting.
17.
Note the SpO2 SAT high and low alarm settings. Raise the SAT low alarm limit above
the actual SpO2 level.
18.
The Warning message SpO2 SAT LOW shall appear on the central alarm display and
a continuous audible alarm shall sound.
19.
Return the SAT low alarm to its previous setting.
20.
Decrease the SAT high alarm limit below the actual SpO2 level.
21.
The caution message SpO2 SAT HIGH shall appear on the central alarm display and
an intermittent audible alarm shall sound.
22.
Return the SAT high alarm to its previous setting.
23.
Remove finger from sensor and set the SpO2 monitor to Standby.
9