INSTALLATION PROCEDURE (continued)

12.

Connect the interface panel cable,
and connect the backplane wire
harnesses and cables. (On earlier
models

without

the

speaker

amplifier board, ensure that the
black wire is connected to the (+)
speaker terminal.)

13.

Record the Novametrix module
serial number and firmware version
number on the PMS form.

14.

Reinstall the monitor box cover and
connect the remote display cable,
BP cuff and BP gauge lines, and
the oximeter sensor cable at the left

side panel of the monitor box.
Replace the disposable reservoir.

15.

Add

the

Operator’s

Instruction

Manual Supplement (P/N 4114884)
to

the

Operator’s

Instruction

binder.

16.

Restore power to the machine.

17.

Perform a PMS procedure on the
machine, including the Novametrix
Monitor calibration check given on
the following pages. Figure 4 shows
the front panel of the Novametrix
TB500B Sensor Simulator, with a
description of its controls.

6

RETURN TO CD-ROM TABLE OF CONTENTS

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS

Содержание Novametrix

Страница 1: ...ld Service Procedure Part Number SP00173 Rev B Date 30 December 1999 1999 Draeger Medical Inc Installation of Novametrix SpO2 Module in Narkomed 4 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 2: ...ng items from the patient interface panel on the left side of the monitor box Remote Display Datagrip cables Pulse Oximeter Sensor cable BP Cuff and BP Gauge lines 3 Remove the disposable reservoir from its holder 4 Remove the screws holding the monitor box cover and carefully lift the cover from the monitor box 5 Disconnect all backplane wire harnesses and cables to the SpO2 assembly and disconne...

Страница 3: ...OX TOP VIEW WITH COVER REMOVED E T SV40438 PULSE OXIMETER ASSEMBLY PULSE OXIMETER ASSEMBLY SPO2 SPEAKER AND AMPLIFIER LATER DESIGN Figure 1 PULSE OXIMETER ASSEMBLY LOCATION 2 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 4: ...BACKPLANE J43 SV40437 EARLIER DESIGNS LATER DESIGN CONNECT WIRE HARNESS TO BACKPLANE J21 AND BACKPLANE J43 CONNECT RIBBON CABLE TO BACKPLANE J12 CONNECT CABLE FROM SpO2 INTERFACE PANEL J1 JP5 JP3 CONNECT CABLE FROM SpO2 INTERFACE PANEL CONNECT CABLE TO BACKPLANE J12 JP5 JP3 Figure 2 NELLCOR PULSE OXIMETER ASSEMBLIES 3 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS...

Страница 5: ...INSTALLATION PROCEDURE continued Figure 3 NOVAMETRIX PULSE OXIMETER MODULE 4 RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Страница 6: ...ard the bottom of the chassis captive screws 7 3B Proceed to Step 8 The following steps apply to the later design SpO2 assembly See Figure 1 and Figure 2 7 1C Remove the four mounting screws securing the SpO2 module to the chassis and remove the module 7 2C Install the Novametrix SpO2 module to the bracket with the hardware supplied in the kit Position the new module in the same manner as the orig...

Страница 7: ...isplay cable BP cuff and BP gauge lines and the oximeter sensor cable at the left side panel of the monitor box Replace the disposable reservoir 15 Add the Operator s Instruction Manual Supplement P N 4114884 to the Operator s Instruction binder 16 Restore power to the machine 17 Perform a PMS procedure on the machine including the Novametrix Monitor calibration check given on the following pages ...

Страница 8: ... simulated Use 86XX for non SuperBrightTM monitors and 87XX for SuperBright 2 RED open push button to simulate red LED failure 3 INFRARED open push button simulates open infrared LED 4 EXTERNAL POWER Jack is used to connect power from an external 9 VDC source When powered externally the simulator s internal battery is disconnected 5 SATURATION selects the simulated saturation value 6 POWER On Off ...

Страница 9: ...y that the monitor indicates 62 2 7 Depress the Open Test RED key and verify that the PULSE LO LED activates a continuous audible alarm is heard and a NO OXI PULSE warning alarm is shown on the central alarm display 8 Depress both Open Test keys together and verify that the monitors SaO2 and OXI ALARMS DISABLE indicators activate a single audible tone is heard and OXI SEN DISC advisory alarm is sh...

Страница 10: ...ssage SpO2 SAT LOW shall appear on the central alarm display and a continuous audible alarm shall sound 19 Return the SAT low alarm to its previous setting 20 Decrease the SAT high alarm limit below the actual SpO2 level 21 The caution message SpO2 SAT HIGH shall appear on the central alarm display and an intermittent audible alarm shall sound 22 Return the SAT high alarm to its previous setting 2...

Страница 11: ...North American Dräger DrägerServiceSM 3122 Commerce Drive Telford PA 18969 215 721 5402 800 543 5047 215 723 5935 Fax RETURN TO CD ROM TABLE OF CONTENTS RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS ...

Отзывы: