INSTALLATION PROCEDURE (continued)
S E N S O R T Y P E
8 6 X X
8 7 X X
R E D
I N F A R E D
E X T E R N A L P O W E R
S V D C 2 5 m A
S I G N A L A T T E N U A T I O N
S A T U R A T I O N
P O W E R
O F F
O N
L O W B A T .
N O V A M E T R I X
M E D I C A L S Y S T E M S I N C .
O P E N T E S T
T B 5 0 0 B
S E N S O R
S I M U L A T O R
O U T P U T
1
2
3
4
5
6
7
8
9
0
1 0 0
9 2
8 2
7 2
6 2
1
2
3
4
5
6
7
8
S P 1 7 3 0 4
Figure 4: TB500B Front Panel
1.
SENSOR TYPE: selects sensor type
to be simulated. Use 86XX for non-
SuperBright
TM
monitors, and 87XX
for SuperBright.
2.
RED: open push-button to simulate
red LED failure.
3.
INFRARED:
open
push-button
simulates open infrared LED.
4.
EXTERNAL POWER: Jack is used
to connect power from an external 9
VDC
source.
When
powered
externally, the simulator’s internal
battery is disconnected.
5.
S A T U R A T I O N :
s e l e c t s
t h e
simulated saturation value.
6.
POWER: On/Off switch; should be
in the Off position when simulator
is not in use.
7.
LOW BAT: LED indicates when the
internal battery is low.
8.
SIGNAL ATTENUATION: varies
the signal strength that the monitor
responds to.
7