WARNING: Upon power up, the monitor will load the last saved alarm
settings. These parameters may not be appropriate for the
current patient’s ventilatory status. The user must review
these alarm parameters to determine their appropriateness
for the current patient and adjust those settings as clinically
necessary.
WARNING: If the NAD Omicron Monitor shows a reading of 127 cm H
2
O
or -127 cm H
2
O, DO NOT CONTINUE TO USE THE UNIT
IN CLINICAL SETTINGS. Remove the unit from use for
servicing.
CAUTION: United States Federal Law restricts this device to sale by or
on the order of the physician.
CAUTION: The NAD Omicron monitor has been designed to work only
with the Narkomed MRI anesthesia machine. Use only NAD
supplied oxygen and flow sensors with this monitor to ensure
accuracy and compatibility with MRI systems.
NOTE:
Due to the sensitivity of the monitor, and while monitoring
within the Respiratory Volume display, a “00” may be
occasionally observed on the Respiratory Volume display. The
presence of the “00” does not influence the accuracy of data or
alarms management, and the display will be updated with
real-time Respiratory Volume data at the completion of the
next expiratory phase.