5-3
Troubleshooting
5
Flow Pressure Data is Missing Throughout Mechanical Ventilation
The probable cause for the flow pressure to be missing throughout the
mechanical ventilation is that the Pressure Low Limit is set too high.
Adjust the Pressure Low Limit to about 4 cm H
2
O below the peak pressure
value.
If a breathing pressure of more than fifty (50) cm H
2
O above the current
peak pressure values was detected, it is necessary to perform a flow
calibration or to power cycle the anesthesia machine to clear the alarm.
WARNING: If the NAD Omicron Monitor shows a reading of 127 cm H
2
0
or -127 cm H
2
O, DO NOT CONTINUE TO USE THE
MONITOR IN THE CLINICAL SETTING. Remove the unit
from use for servicing.