6-4
6
Specifications
Flow
The following is Flow specification information:
Transducer
Type
Pitot tube incorporated into reusable flow tube.
Accuracy
Better than 5.0% Full Scale Output if O
2
sensor employed
Displayed Information
The following is monitor display specification information:
Pressure
Range
•
Operating Range: -10 to +120 cm H
2
O
•
Display Range:
–
-10 to + 120 cm H
2
O. Peak
–
-10 to + 120 cm H
2
O. Mean
–
-10 to + 120 cm H
2
O. PEEP
•
Resolution:
–
1.0 cm H
2
O on bar graph display
–
0.1 cm H
2
O on numeric display
Respiratory
Rate
•
Operating Range: 2 to 400 BPM
•
Display Range: 2 to 400 BPM
•
Resolution: 0.1 BPM on numeric display
FIO
2
•
Operating Range: 18% to 100%
•
DisplayRange: 0% to 100%
•
Resolution: 0.1% on numeric display
•
(
O
2
SENSOR
in case of O
2
Sensor Failure)