3-9
Operating Instructions
3
4. After the pressure calibration is complete, the monitor will display
PRESS OK
and
0000
.
5. Within three seconds, the monitor will display
FLOW OK
and
0000,
indicating that it is establishing the zero set points for the two flow
channels (Plus and Minus Flow).
6. After the tidal volume calibration is complete, the monitor will display
TIDAL VOL
and
----.
7. Finishing this sequence indicates that the monitor has successfully
completed both pressure and tidal volume zero calibrations and will
achieve pressure and volume accuracy specifications.