Instruction Manual Dräger MSI EM2 EM200plus-i
36
17.2 Error messages
Error message
Error cause
Remedy
O2 sensor
probe has been in flue
sensor defect
calibrate again with fresh air
service
CO sensor
probe has been in flue
sensor defect
calibrate again with fresh air
service
NO sensor
NO sensor has been without current
others like CO sensor
check function, load battery
see CO sensor
pressure sensor
sensor with pressure during calibration
sensor defect
calibrate again without pressure
service
Error message
Error reason; remedy
system temperature
operating temperature range exceeded;
vary temperature, service
junction temperature
operating temperature range exceeded;
vary temperature, service
battery temperature
operating temperature range exceeded;
vary temperature, service
battery voltage
voltage not between 4.4 V and 6.5 V; replace battery (service)
battery current
battery current for charging or operating too high; service
battery error
battery manager data error; charge battery
set clock
clock lost data; charge battery, set clock
gas flow
gas flow out of operating range; check gas duct, service
options
data record options error; factory service
settings
wrong settings; check and change settings
calibration data
error in calibration data record; service
data memory
error in memory data record; service
next service
next service data error; service
fuel table
fuel table error; reload fuel table with PC program "PC200P"
display table
display table error; reload display table with PC program "PC200P"
printer table
printer table error; reload printer table with PC program "PC200P"
system configuration
system configuration error, service