24
Dräger X-dock 5300/6300/6600
Basics
4.6
Configuring the test gas inlet
To configure a gas inlet:
1. Select
>
Station gas configuration
.
An overview of the test gas connections is displayed.
2. Select desired gas inlet.
The configuration menu will appear.
Default setting for the bump test tolerance: 20 % (5 % with O
2
)
When using a Dräger gas cylinder:
1. Enter the part no. of the Dräger gas cylinder.
All the necessary details for the configuration will be filled
in automatically. The batch number and the expiry date can
be entered manually in addition.
2. If necessary select
Further options
and , to reset the
gas cylinder level indicator.
3. If required, configure other gas inlets in the same way.
When using a gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with .
Delete the current test gas component with .
2. Select gas.
3. Enter the gas concentration.
4. Select gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part no. of the gas cylinder
Batch number of test gas cylinder
Expiry date of the gas cylinder
7. If required, enter
Further options
.
8. For more information see X-dock 5300/6300/6600
Technical Manual.
WARNING
The gas concentrations entered must be identical to
the specifications on the gas cylinder used. Incorrect
details will result in faulty measurement results.
NOTICE
When the test gas concentration is changed, the
corresponding test gas inlet must be reconfigured.
Select the test gas inlet
►
Select the test gas
inlet
►
Enter part no.
►
◄
Enter batch no.
Enter use-by date
►
Select test gas
►
◄
Create Test gas
component
◄
Delete configuration
Enter concentration
►
◄
Select unit
◄
Set bump test
tolerance
Further options
►
!
i
i
NOTICE
Upon entering the part no. of a Dräger gas cylinder, a
gas cylinder level indicator is displayed automatically,
unless this function was deactivated before (see
NOTICE
The values entered automatically must be matched
with those indicated on the test gas cylinder. If these
values are not identical, the value indicated on the gas
cylinder is to be considered and a manual correction of
the values must be made in the station.
NOTICE
Deleting all test gas components deletes all
information of the test gas inlet.
i
i
i
i
i
i