PMC PROCEDURE (continued)
6.17.6
Adjust the O
2
flow to 3 l/min.
6.17.7
Is the tidal volume on the test volumeter greater than 250 ml? ___(Y)
6.17.8
Adjust the tidal volume to the 100 ml mark on the pediatric bellows
assembly.
6.17.9
Is the tidal volume on the test volumeter within 65 to 135 ml? ___ (Y)
6.17.10
Verify that with the Pediatric Bellows Fine Flow Control turned fully
counter-clockwise the bellows does not collapse during inspiration.
Readjust the knob to the fully clockwise position.
6.17.11
Adjust the O
2
flow to 10 l/min.
6.17.12
Is the PEEP displayed on the monitor 0 to 3 cm H
2
O? ___ (Y)
6.17.13
Close the Oxygen flow control valve.
6.17.14
Remove the ventilator hose from the VENTILATOR HOSE terminal.
6.17.15
Attach a test terminal (P/N 4104389) to the bellows assembly ventilator
hose terminal.
6.17.16
Connect a 0-250 ml/min. flowmeter (P/N S000081) to the test terminal.
6.17.17
Set the FREQUENCY to 1 BPM.
6.17.18
Set the I:E RATIO to 1:1.
6.17.19
Set the Inspiratory Flow to MAX.
6.17.20
Is the drive gas leakage indicated during the inspiratory phase 0 to 50
ml/min? ___ (Y)
6.17.21
Remove the test equipment and reattach the ventilator hose to the
VENTILATOR HOSE terminal.
6.17.22
Set AUTO/BAG valve to BAG.
6.17.23
Press the Alarms All Standby key.
6.18 BELLOWS PEDIATRIC INTERNAL - If applicable
6.18.1
Remove the breathing bag from the Y-piece and replace it with a test
lung (P/N 4115128).
6.18.2
Press the O
2
Flush momentarily to inflate the bellows.
6.18.3
Does the bellows remain fully inflated during the expiratory pause
phase? ___ (Y)
6.18.4
Set the ventilator frequency to 20 BPM.
6.18.5
Set the Oxygen flow to 3 liters.
6.18.6
Is the tidal volume on the test volumeter greater than 250 ml? ___ (Y)
6.18.7
Adjust the inspiratory flow and Pressure limit control if applicable
control until the bellows collapses to the 100 ml mark on the pediatric
bellows assembly.
6.18.8
Is the tidal volume on the test volumeter 65 to 35 ml? ___ (Y)
(
9
)
(
9
)
(
9
)
(
9
)
(
9
)
(
9
)
(
9
)
Содержание Narkomed GS
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...