Setup and Installation
3-16
Narkomed GS Setup and Installation Manual
Draeger Medical, Inc.
Pole Mount
The Bain Circuit Adapter that mounts on the absorber pole may be
supplied with an optional positive end-expiratory pressure (PEEP) valve.
To install a pole-mounted Bain Circuit Adapter, with or without a PEEP
valve:
1. Loosen the set screw on the absorber pole and remove the absorber
system (if the machine is so equipped).
2. Slip the Bain Circuit Adapter mounting stud into the absorber pole.
3. Tighten the set screw on the absorber pole to lock the Bain Circuit
Adapter in place.
OXYGEN
SENSOR
BAIN
CIRCUIT
BREATHING
BAG MOUNT
19 MM SCAVENGER HOSE
(P/N 9995230)
(CONNECTS TO 19MM
TERMINAL ON SCAVENGER)
22MM HOSE
(P/N 9995123)
(CONNECTS TO
VENTILATOR HOSE
TERMINAL ON
VENTILATOR
BELLOWS ASSEMBLY)
BAIN
CIRCUIT
HOSE BARB
FOR FRESH
GAS HOSE
MOUNTING
STUD
FRESH GAS
COMMON
OUTLET
FRESH GAS
HOSE
SU14719A
BREATHING BAG
SET SCREW
BREATHING
PRESSURE
PILOT LINE
Содержание Narkomed GS
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...