Setup and Installation
3
Draeger Medical, Inc.
Narkomed GS Setup and Installation Manual
3-11
For Y-piece
Monitoring
For breathing pressure monitoring at the patient Y-piece, install the long
pilot line (which has a quick-connect fitting on one end and a Luer type
fitting on the other end) as follows:
1. Connect the quick-connect fitting on the pilot line to the BREATHING
PRESSURE interface underneath the rear panel of the ventilator box.
NOTE:
The quick-connect fitting on the absorber is self-closing. You can
leave it unused when installing the long pilot line for
Y-piece monitoring.
2. Connect the Luer fitting on the other end of the pilot line to a Luer
fitting on either the patient Y-piece or a 15 mm adapter on the patient
side of the Y-piece.
3. Using the four plastic hose clips attached to the pilot line, mount the
pilot line on either of the breathing hoses leading to the Y-piece.
SU00604A
REAR PANEL OF
VENTILATOR BOX
2
VOLUME
SENSOR
OXYGEN
SENSOR
BREATHING
PRESSURE
AUTO/BAG
SELECTOR
MALE
LUER-LOCK
FITTING
QUICK-CONNECT
FITTING
LONG
PILOT LINE
(P/N 4108528)
PLASTIC
HOSE CLIPS
Содержание Narkomed GS
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...