Setup and Installation
3
Draeger Medical, Inc.
Narkomed GS Setup and Installation Manual
3-5
Installing the
Vaporizers
(Removable
Vapor Mount
System)
The removable vapor mount system will accommodate Dräger 19.3 and
Dräger 2000 vaporizers.
If the customer does not specify the vaporizer positions, install the
vaporizers in the following order, as viewed from the front of the machine:
Left position:
Halothane
Right position:
Isoflurane
Install the vaporizers as follows:
1. Examine the ports on the vaporizer manifold and verify that each port
has an O-ring.
2. Move the selector lever to the right.
3. Place the vaporizer down over the left manifold ports, and turn its
locking lever 90
°
clockwise to secure the vaporizer to the manifold.
4. Move the selector lever to the left.
5. Place the second vaporizer down over the right manifold ports, and
turn its locking lever 90
°
clockwise to secure the vaporizer to the
manifold.
6. Test the interlock for proper operation. Verify that only one vaporizer
at a time can be turned on. Return all vaporizers to their Zero position
at completion of the test.
Содержание Narkomed GS
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Страница 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...