Things to Know Before Taking Measurement

To assure accurate measurement, please observe the following:

Make sure that electrodes are directly in contact with your skin.

If your skin or hands are dry, moisten your skin with damp cloth before taking  measurement.

 

If the electrodes are dirty, use a soft cloth or a cotton swab moistened with disinfectant alcohol to

       clean dirt from the surface.

During the measurement, do not touch your body with the hand with which you are taking the

       measurement.

Make sure there is no skin contact between your right and left hand (measuring method C), or

       between your hand and chest (measuring methods A/B). Otherwise, the measurement cannot be
       taken correctly.

Ideally perform the measurement either sitting down or lying down. Do not stand up. 

Do not move or talk during the measurement and hold the device still. Movements of any kind will

       falsify the measurements.

6.  TAKING MEASUREMENT

Operation Procedure

There are three different methods of taking the measurement. Start with measurement method A, “right 
index  nger-chest.”  If  this  method  does  not  deliver  measurements,  or  delivers  only  unstable 
measurements  (“EE”  is    frequently  displayed),  switch  to  method  B “left  index 

chest”  and,  if 

necessary, to method C”left hand-right hand.”

The most suitable method/ procedure depends on the heart 

(shape of the heart) of each 

individual user. If it is not possible to take stable measurements using a given measurement method, this 
could have a harmless cause such as the shape of the heart. However, the cause may also be an illness or 
disease.

Measurement method C offers maximum comfort, but offers a much lower measurement stability than 
methods A or B.

Press the on/ off button for about three seconds to turn the device on. The device shows the full display.

WARNING

If segments are missing, stop using the device and immediately contact customer services. To test whether 

the full display is completely displayed, hold the On/Off button when switching on the device.

Measurement Methods

Measurement

method A 

chest”

(approximately corresponds to “lead 2”)

Top Electrodes

Place y

wo upper electrodes on the device 

and hold the device

You can determine the correct position for placing the lower electrode of 
the device against your chest using the following methods:

•  Draw an imaginary line from the front of your armpit downwards. At the 
same time, draw an imaginary line going up 10 cm from the lowest rib on 
the  left  side  of  your  body.  Place  the  lower  electrode  of  the  device  at  the 
point where these two lines meet.

5 cm

•    Draw  an  imaginary  line  from  the  bottom  of  the  centre  of  your  chest 

(sternum) to the left. At the same time, draw an imaginary line from the front 

of  your  armpit  downwards.  Place  the  lower  electrode  of  the  device  at  the 

point where these two lines meet.   

9

8

Содержание ECG PEN

Страница 1: ...e by providing with the most trusted and innovative health care and wellness products Being a renowned global leader in health care products Dr Trust ensures that our technically efficient team works dynamically and tirelessly to provide the best of the medical devices to our clients The products that we havetoofferaresuitablydesignedforuseathomes laboratoriesandhospitals Our ground breaking solut...

Страница 2: ...ement 6 Taking Measurement 7 Result Display 8 iBP Monitor APP 3 3 4 5 7 8 13 14 9 Cleaning and Maintenance 10 Messages and Trouble Shooting tions 12 EMC Information 13 Customer Support 16 17 18 19 25 1 SYMBOL DEFINITIONS 2 GENERAL NOTES 1 ThisisanElectrocardiogram ECG devicethatprovidesreferentialdatathatcouldbetakeninto considerationbymedicalprofessionals 2 TheappliedpartisECG 3 Selfdiagnosisortr...

Страница 3: ...Donotstore exposethedeviceinthefollowingenvironments Locationsexposedtodirect sunshine hightemperatureorhumidity heavydirt nearwateror e andunderhighelectro magnetic Storethedeviceinsafeplacetoavoidsmallpartsswallowedbychildrenunderanycircumstances Static electricity may cause damage to the electronic components to preventelectrostaticdamagetothe device grounding yourself to discharge the static e...

Страница 4: ... Digits Display a Shows the sequence number of the measurement results stored in memory b Shows the 30 second count down during measurement 5 Time Shows date and time 6 Bluetooth Connection Displays the status of Bluetooth connection 5 PREPARATION BEFORE TAKING MEASUREMENT To charge your ECG Pen please follow the steps below 1 Remove the battery cover by inserting the tip of y er 2 Follow the posi...

Страница 5: ...of each individualuser Ifitisnotpossibletotakestablemeasurementsusingagivenmeasurementmethod this could have a harmless cause such as the shape of the heart However the cause may also be an illness or disease Measurement method C offers maximum comfort but offers a much lower measurement stability than methodsAorB Presstheon offbuttonforaboutthreesecondstoturnthedeviceon Thedeviceshowsthefulldispl...

Страница 6: ... against your chest until you hear or feel a click Important Do not exert excessive pressure on the electrodes with y Important Do not pr our skin Measurement method C left hand right hand approximately corresponds to lead 1 Place y wo upper electrodes on the device our left hand on the lower electrode Press the lower electrode until you hear or feel a click Important Do not pr our skin Please not...

Страница 7: ...lding ECG Pen in the wrong direction 7 Result Display Normal Result Slow Heart Rate 60 bpm Fast Heart Rate 100 bpm Paused Heart beat pause 2 s Irregular Rhythm Disrupted Rhythm Abnormal Waveform Altered Waveform Please note that if the displayed heart rate is it indicates ECG signals are unstable or weak Pleaseretakethemeasurement Measurement Results The possible results are shown below 13 12 ...

Страница 8: ...dGoogleplaystoresforandroidusers Downloading the app You can connect IBP monitor app with your Dr Trust ECG Pen devices using Bluetooth 4 0 It helps in providing real time display of ECG and blood pressure measurements The app also automatically stores recordsonsmart phoneortabletforfurtherdiagnosis QR code scanning Application Interface Measure Using the measure feature ECG and blood pressure can...

Страница 9: ...e Shown Possible Cause Suggestion Measurement result shows t Rate Unstable or weak ECG signals Please retake the measurement following the manual s instruction Measurement result shows EE Failure analysis resulted from either low contact pressure or h i g h n o i s e l e v e l d u r i n g measurement Please retake the measurement following the manual s instruction Problem Possible Cause Suggestion...

Страница 10: ...nitor APP Problems of data exchanging for tion If the device is discovered successfully there will be a blank on the box then click on the blank to manually link the device with APP and display the device ID 11 SPECIFICATIONS 12 EMC INFORMATION 19 18 ...

Страница 11: ...21 20 ...

Страница 12: ...bserved to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating theDr TrustECGPen Recommended separation distances between portable and mobile RF communications equipment and the Dr Trust ECG Pen The Dr Trust ECG Pen is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled T...

Страница 13: ...umoutputpowerratingofthetransmitterinwatts W accordingto thetransmittermanufacturer NOTE1At80MHzand800MHz theseparationdistanceforthehigherfrequencyrangeapplies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorptionandr tionfromstructures objectsandpeople COPYRIGHT 2018 DR TRUST ALL RIGHTS RESERVED www drtrustusa com dr drtrustusa com drtrust ...

Отзывы: