![Dr.Thiedig Digox 6 Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/dr-thiedig/digox-6/digox-6_operating-manual_2526274046.webp)
46
8 Special functions
The special functions of the
Digox 6
allow to poll
device states or to make settings.
8.1 Status
display
Figure 8.1
Use
and the menu item STATUS for the status
display. Here, the following information is
provided:
•
Ks
factor
refer to section 4.4
•
Kf
factor
refer to section 4.3
•
zero
current
refer to section 4.7
•
Medium
here, the medium setting is displayed,
the default setting for the
Digox 6
is
Pilsner
•
measuring
range
here the setting of the recorder output is
displayed.
NOTE:
The display of the measuring range does not refer
to the measuring range used by the
Digox 6
. Here
the current setting for the measuring range of the
recorder output is displayed.
8.2 Selecting recorder measuring
ranges and output currents
1. In the main menu, press
until the menu
item MEASURING RANGES is highlighted.
Press
to open the menu MEASURING
RANGES.
Figure 8.2
2. Select the recorder output and confirm with
. The current measuring range is
highlighted in the menu opened. Use
and
to set the desired measuring range and
to confirm. To save the setting, highlight
the menu RETURN and press
.
Figure 8.3
The following measuring ranges can be
selected for the recorder outputs:
•
concentration 1,000mg/l, 20,000mg/l,
automatic identification of the measuring
range
NOTE:
The concentration ranges are set in the works, if
required by the customer, they can be adapted by
Dr. Co service personnel.
3. After saving, use
and
in the
subsequent menu to set the output current
and save the value with
.
Содержание Digox 6
Страница 1: ...Operating Manual Oxygen Analyser...
Страница 7: ...7...
Страница 11: ...11 Figure 1 3 1 Display 2 Sample supply In Figure 1 4 1 6 1 2...
Страница 13: ...13 Notes...
Страница 45: ...45 Notes...
Страница 57: ...57 10 5 Menu overview...
Страница 59: ...59 Notes...
Страница 61: ...61 Touch Memory System 9 Troubleshooting 42 U Unit description 8 Unit extensions 56 Z Zero current 19 26...