4.10. Positioning
Use the sterile handle
14
or the handle rails
13
to
position the lamp.
Use the handle rail to position the lights before the
operation.
Use the sterile handle for positioning during the oper-
ation. This handle can be removed for sterilization.
5. Cleaning
5.1. Sterilisable handle
At delivery the lamp is equipped with the handle
sleeve
14
. The handle sleeve is removable and
sterilisable. Before using the first time and before
every use the handle sleeve must be cleaned, disin-
fected and sterilised.
The handle sleeve must be removed for sterilisation:
To remove press the lock
V
and pull off the steri-
lisable handle sleeve
14
while keeping the lock
pressed.
To attach, push on and slightly twist the handle
until the lock
V
engages securely.
Handles often become unsterile during an
operation. Therefore always keep additional handles
available for exchange.
Cleaning / disinfection and sterilisation
Basics
Efficient cleaning / disinfection is an essential requirement for effective sterilisation of the handle.
Within the scope of responsibility for the sterility of the products it should be noted that only sufficiently
validated equipment and product specific processes are used for cleaning / disinfection and that the vali-
dated parameters are complied with in every cycle.
In addition, the hospital / clinic hygiene regulations must be observed.