Mach LED 150F / 150 / 150FP
59320002
Edition 12
08.08.2018 / Bak We page 13/26
Inspection / durability
The sterilisable handle sleeve must be disposed
after 1000 sterilisation cycles or at the latest after
2 years and replaced with a new one.
The year of manufacture can be determined with
the help of a stamping on the inner side of the
handle sleeve (like shown in the photo). The
stamping in the photo shows the number 12,
which stands for the year 2012.
4.2 Lamp housing, protective disk
and support system
The Dr. Mach light system has a high-quality
surface, which can be cleaned with conventional
cleaning agents.
The protective disk
8
is made of a high-quality
plastic. Pay attention to the following during
cleaning:
-
Never wipe over the protective disk
8
with a dry
cloth (always clean with a wet cloth).
-
Wipe the protective disk
8
after cleaning with an
antistatic, non-fluffy cloth.
8