Mach LED 150F / 150 / 150FP
59320002
Edition 12
08.08.2018 / Bak We page 24/26
Table 5 – Input a.c power PORT (1 of 2)
Phenomenon
Basic EMC
standard
Immunity test levels
Professional healthcare
facility environment
Home healthcare
environment
Electrostatic fast transients
bursts
a) l) o)
IEC 61000-4-4
± 2 kV
100kHz repetition frequency
Surges
a) b) j) k) o)
Line – to- line
IEC 61000-4-5
± 0,5 kV, ± 1 kV
Surges
a) b) j) o)
Line – to- ground
IEC 61000-4-5
± 0,5 kV, ± 1 kV, ± 2 kV
Conducted disturbances
induced by RF fields
c) d) o)
IEC 61000-4-6
3 V
m)
0,15 MHz - 80 MHz
6 V
m)
in ISM bands
between 0,15 MHz and
80MHz
n)
80 % AM at 1 kHz
e)
3 V
m)
0,15 MHz - 80 MHz
6 V
m)
in ISM and amateur radio
bands between 0,15 MHz and 80MHz
n)
80 % AM at 1 kHz
c)
Voltage dips
f) p) r)
IEC 61000-4-11
0 %
U
T
; 0,5 cycle
g)
At 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
q)
0 %
U
T
; 1 cycle
and
70 %
U
T;
25/30 cycles
h)
Single phase: at 0°
Voltage interruptions
f) i) o) r)
IEC 61000-4-11
0 %
U
T;
25/30 cycles
h)
a)
The test may be performed at any one power input voltage within the ME EQUIPMENT OR ME SYSTEM RATED voltage range. If
the ME EQUIMENT OR ME SYSTEM is tested at one power input voltage, it is not necessary to re-test at additional voltages.
b)
All ME EQUIPMENT and ME SYSTEM cables are attached during the test.
c)
Calibration for current injection clamps shall be performed in a 150.Ω system.
d)
If the frequency stepping skips over an ISM or amateur band, as applicable, an additional test frequency shall be used in the ISM
or amateur radio band. This applies to each ISM and amateur radio band within the specified frequency range.
e)
Testing may be performed at other modulation frequencies identified by the RISK MANAGEMENT PROCESS.
f)
ME EQUIPMENT and ME SYSTEMS with a d.c. power input intended for use with a.c.-to-d.c. converters shall be tested using a
converter that meets the specifications of the MANUFACTURER of the ME EQUIPMENT or ME SYSTEM. The IMMUNITY TEST
LEVELS are applied to the a.c. power input of the converter.
g)
Applicable only to ME EQUIPMENT and ME SYSTEMS connected to single-phase a.c. mains.
h)
E.g. 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.
i)
ME EQUIPMENT and ME SYSTEMS with RATED input current greater than 16 A / phase shall be interrupted once for 250/300
cycles at any angle and at all phases at the same time (if applicable). ME EQUIPMENT and ME SYSTEMS with battery backup
shall resume line power operation after the test. For ME EQUIPMENT and ME SYSTEMS with RATED input current not exceed-
ing 16 A, all phases shall be interrupted simultaneously.
j)
ME EQUIPMENT and ME SYSTEMS that do not have a surge protection device in the primary power circuit may be tested only at
± 2 kV line(s) to earth and ± 1 kV line(s) to line(s).
k)
Not applicable to CLASS II ME EQUIPMENT and ME SYSTEMS.