RECOMMENDATIONS FOR CARRYING OUT NEBULIZING THERAPY:
•
Sit down in a comfortable and upright position.
•
If a mask (adult or pediatric) is used, be sure that it securely adheres to the face. Mask has holes for the air flow during the expiration.
•
Breathe easy and have a short pause at the end of each breath.
•
Therapeutic application ends when the medicine came out in a discontinuous mode (spluttering).
•
When using the mouthpiece, do not tilt the medicine cup over 45°.
Changing the Air Filter
The filter has to be substituted after 30 hours of use or when it becomes grey. Spair air filter and its container
are on the figure H. For filter replacing, carefully raise the lid using a flat screwdriver as shown in the picture I.
Remove the filter and fit in a new filter (9). Fit the filter carrier into its housing making sure it is attached securely.
UNIT MAINTENANCE
Cleaning the Unit
After each therapy remove the device from the socket and clean the device and the accessories. Attention: use only the accessories supplied with the product,
other accessories can lead to ineffective therapy.
CLEANING OF THE MOUTHPIECE
Use drinking water to carefully clean these accessories. They can be sterilized by boiling for 5-10 minutes into the water or by immersion in a chemical disin
-
fectant (oxidizing electrolytic chloride), following product instructions and rinsing off with drinking water. Leave to dry naturally in a clean place.
CLEANING OF THE MEDICINE CUP
After use, open the two parts, wash with soap and hot water and rinse it carefully. Remove excess of water with a soft, clean and dry cloth and leave to dry
naturally in a clean place. To sterilize, cover the medication container with a cold sterilizing solution, following the instructions on the sterilizer pack. It is rec
-
ommended to use one personal medication container for each patient, to be use for 6 months or 120 treatments. Do not use boiling water to clean it. Carefully
wash masks and hoses with disinfectant solution only.
CONTAMINATION BY MICROBES
In case of pathologies of microbial or viral origin, where infection is possible, the device and its accessories should not be shared with other patients.
Troubleshooting
PROBLEM
ACTION
The device is not turning on.
•
Make sure that the plug has been correctly inserted in the main
supply.
•
Make sure that the switch is in “ON” position.
The device is turned on but it’s not nebulizing.
•
Make sure to have inserted the nozzle (7) into the medicine cup.
•
Make sure that the air tube is not bended or crushed (8).
•
Make sure that the air intake is not obstructed and the air filter is not
dirty. Replace the air filter with the new one if necessary (9).
•
Make sure that the medicine cup has been filled in with medicine.
NOTE:
If the device is not functioning correctly in spite of inspections effected, contact the ТМ Dr. Frei
®
service center in your country.
TECHNICAL SPECIFICATIONS
Compressor nebulizer machine. Complies with the requirements of European Standard EN 13544-1. Piston rotary presser, without lubrification. Medical device
of class IIA according to Directive 93/42/EEC. Medicine Linear Cup.
Model
Turbo Base
Adapter input
100 - 240 V ~ , 50 / 60 Hz, max 0.5 A
Adapter output
12.0 V
1 A
Main unit Input
DC 12V, 1A
Noise Level
≤ 55 dB
H
I
Содержание TURBO CAR
Страница 8: ...8 36...
Страница 13: ...UA 13 16 17 I I I 17 i 36...
Страница 16: ...16 10 C 40 C 85 860 1060 2 II 2000 5 110 x 110 x 62 2000 8 2460 BF ON OFF IP 21 12...
Страница 23: ...BG 23 36 c Wi Fi...
Страница 26: ...26 10 C 40 C 85 860 1060 2 II 2000 5 110 x 110 x 62 2000 8 2460 93 42 BF ON OFF IP 21 12...
Страница 33: ......