43
During transportation, the instrument should be protected from dust and precipitation.
It is
forbidden to tilt and turn over the instrument!
Prior to the shipment, the instrument should
be secured to ensure its stable position and exclude any displacement and shocks.
Instrument can be shipped by any kinds of transport and it has to be carried in the
manufacturer’s transport package and treated with accordance to manipulation signs applied
on the outer surface of transport package. Acceptable environmental conditions for
transportation are the following: temperature from - 50°C to + 50°C and relative air humidity
up to 80 % at + 6°C.
Note!
The instrument contains precision mechanical parts. To avoid their damage during
handling and shipping, the requirements of manipulation signs applied on package should be
strictly followed.
Attention!
The instrument has to be shipped with the closed thermal block only.
7. Recycling
7.1 Recycling of medical devices has to be carried out in accordance with the classification,
rules for the collection, use, disinfection, storage, transportation, accounting and disposal of
medical waste established by the authorized federal executive authority.
7.2 Recycling of the instrument has to be carried out by special companies, which have the
appropriate license, using specially equipped places in accordance with the requirements
stipulated by existing Federal laws and in compliance with the mandatory requirements for
environmental protection in accordance with the local regulations.
8. EMS declaration
The instrument complies with the immunity and electromagnetic emission requirements of
EN 61326-2-6: 2013.
The instrument is intended for use in the electromagnetic environment described below:
-
The device was designed and tested in accordance with the requirements of
CISPR 11 and it is suitable for use in all rooms, including ones, which are
directly connected to the public low-voltage mains network.
-
The quality of the electricity supply has to comply with typical conditions of
use in commercial institutions or hospitals.
-
Power frequency of magnetic fields has to be at a level corresponded with
typical conditions of use in commercial institutions or hospitals.
-
The instrument uses radiofrequency energy exclusively for its internal
function. The level of radiofrequency emission is very low and does not lead to
disturbances in the operation of electronic equipment located close to it.
-
It is forbidden to use the instrument near sources of strong electromagnetic
radiation that could interfere its normal functioning.
Notes:
1.
The manufacturer is responsible for providing the consumer information about the
electromagnetic compatibility of the instrument.
2.
The customer is responsible for maintaining electromagnetic environment, which
ensures instrument normal operation in accordance with its intended purpose.