38
The same message will appear at the instrument’s display as well. When warming up is
completed, the instrument will automatically proceed to the RUN execution mode.
In case of necessity of instrument’s autonomous functioning after the RUN completion and its
automatic turning to sleep mode perform the following:
1. In
Running
window activate the option - turn off device after the RUN
2. Close the Real Time PCR software by clicking
in the upper right corner of its
window.
3. When RUN is completed the instrument will switch to the sleep mode, what is confirmed
by blinking diode on the front panel of the instrument.
Note!
The instrument will proceed to the sleep mode only if “
Turn off device after the run”
option is activated and the executed amplification program finishes as
Holding
mode.
4.6 Amplification program running
Running of the amplification program is controlled via RealTime_PCR software.
The following options could be done during its execution:
pause running;
stop running;
add any number of cycles in a current block or skip execution of the remaining cycles
and proceed to the execution of the next temperature block;
edit the log of optical measurement;
to check optical measurement data of the selected tubes in the form of 3D histogram in
the
Distribution of optical measurements over the plate
window by clicking
(Fig. 28).