![DMT UHSAS-0.055 Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/dmt/uhsas-0-055/uhsas-0-055_manual_2514260010.webp)
UhsasManualRevE060405DMT.doc,10 / 34
Chapter 4 - Theory of operations
Instrument subsystems
The UHSAS is an optical-scattering, laser-based aerosol particle spectrometer system for
accurately and precisely sizing particles in the range from 0.055
µ
m to 1.0
µ
m diameter.
It uses fully user-specified size binning of up to 100 channels anywhere within its size
range.
The spectrometer instrument consists of 5 general subsystems, described in this chapter.
1)
the main optical subsystem responsible for generating the laser light, detecting the
scattering from the particles and providing a mechanical enclosure for the optical
system and for delivery of the sample aerosol
2)
the flow system for bringing the sample aerosol through the optical interaction
region, including flow control and measurement
3)
the analog electronics system for amplifying and processing the particle signals
4)
the digital electronics system for analyzing particle signals, binning signals
according to user-specified bin mappings and generating a histogram of number
of particles in the specified bins, and for communicating with the PC and system
monitor/control functions
5)
an onboard PC running Windows and a specialized application GUI for
instrument control, setup and data reporting and collection
Figure: Block diagram of optical particle spectrometer
1) Optical system
Digital electronics (4)
PC & Display (5)
Monitor &
control
Histogramming
functions
Analog signal
processing (3)
Optical
system (1)
(2) Flow
system
Flow
control &
monitor
laser
control &
monitor
Flow
Analog monitor
Peak-
held
analog
particle
signal
Particle
photo-
signal
ADC