38
3.7 SETTING MENU
The Settings Menu for the 114P system contains several sub-
menus. These sub- menus include:
• Force Calibration
• Pressure Calibration
• pH Calibration (optional)
• Select Analog Output
• Interface Settings
• Select 114P/112PP
The Settings Menu for the 112PP system contains several
sub-menus. These sub-menus include:
• Force Calibration
• Pressure Calibration
• pH Calibration (optional)
• Select Analog Output
• Interface Settings
• Pulse Rate
• Cleaning Menu
• Select 114P/112PP
3.7.1 FORCE CALIBRATION MENU
Prior to shipping, the Pressure Myograph System has gone through two days of continuous testing,
including final force calibrations. However, DMT recommends that new force calibrations are performed
before starting to use the Pressure Myograph System for the first time. The calibration procedure needs
to be performed to calibrate the system. Before starting the force calibration, the following is performed.
Please see figure 3.1 and 3.2 for Pressure Myograph – 114P or Pulsatile Pressure Myograph - 112PP,
respectively.