5. Trouble Shooting 

 

Trouble Shooting Chart

 

Problem

 

Possible Cause(s)

 

Solution

 

Motor will 

not Start 

when 

Turned 

ON

 

Check all wires for loose 

or broken connections. 

Check fuse. Check low 

voltage of main power  

If voltage is not present, replace the 

transformer or fuse.  

Test solenoid valve by 

loosening  brass nut on right 

side. If water flows out with 

power on, turn power off. 

Water flow should stop. 

Caution: Do not 

operate the 

pump for an extended time, 

because running the pump 

without water could cause 

internal  damage. 

If water flow does not stop with  power 

off, replace the solenoid  valve. 

Motor 

stops or will 

not start 

Circuit breaker, main cut-

off  and  low-  voltage 

operatory switches are in 

the 

OFF 

position. 

Place switches in 

ON 

position. 

Unit is not plugged  in. 

Plug in unit. 

Loose or broken wires. 

Tighten, repair or replace  wires. 

Motor hums indicating a bad 

capacitor. 

Replace capacitor. 

Tight or noisy motor. 

Check bearings. Make sure pump is 

properly shimmed to the motor. Or, 

remove any debris in the  pump. 

 

 

 

 

 

 

 

 

 

Содержание Mega Power I

Страница 1: ...s intended use as described in this manual Improper use or failure to do so could result in personal injury or property damage This device is to only be used by dental staff or dentist in dental pract...

Страница 2: ...m Page 9 5 Trouble Shooting Page 11 6 PartList Page 13 7 Warranty Page 14 1 Introduction of Intended Use A This Suction unit is intended for use of dental suction to suck contaminated water from denta...

Страница 3: ...3 2 Product Structure 1 Explosive view of product...

Страница 4: ...llen into water immediately unplug the unit 2 Never operate the unit if it has any damaged parts including power cord and or if it has been dropped or submersed inwater 3 The unit should not be used w...

Страница 5: ...V Mega Power I 115V Model name MP1 MP1 115 1 Voltage AC 210V 235V AC 110V 120V 2 Hz 50 60Hz 3 Amp 6 Amp 12 Amp 4 H P 1 HP 0 75kw 5 Capacitor 450V 40uF 250V 70uF 6 Max Negative Pressure 450 550mbar 4 6...

Страница 6: ...Mega Power Liquid Ring Vacuum Pump 6 4 Installation Instruction A Plumbing drawing...

Страница 7: ...er switch and check any leakage of pump vacuum level 3 Maintenance Daily Maintenance Clean Vacuum Piping System Clean daily The vacuum piping should be cleaned using a non foaming vacuum system cleane...

Страница 8: ...ter inlet strainer if necessary or every 1 year The water inlet strainer has small wire mesh inside to remove rust Use large crescent wrench to remove nut With air gun clean mesh inside of strainer an...

Страница 9: ...9 4 Wiring Diagram Circuit Board Remote Switch cable Circuit Breaker Capacitor Fuse Water Inlet Strainer Solenoid Valve...

Страница 10: ...Mega Power Liquid Ring Vacuum Pump 10...

Страница 11: ...ution Do not operatethe pump for an extended time becauserunning the pump without water couldcause internal damage If water flow does not stop with power off replace the solenoid valve Motor stopsor w...

Страница 12: ...alve and if possible remove debris Otherwise thevalve should be replaced Pump is worn out Check the pump inside and replace it Insufficient water supply Check water supply is sufficient Too excessive...

Страница 13: ...13 Part List...

Страница 14: ...luding but not limited to those for use of any products loss of time inconvenience lost profit labor charges or other incidental or consequential damages with respect to persons business or property w...

Страница 15: ...Customer designs assemblies constructs or manufacturers This warranty can be modified only by authorized DMEGA personnel by signing a specific written description of any modifications MODEL Mega Power...

Страница 16: ...Address 122 Podowon Ro 861 2 Sincheon Dong Siheung City Gyunggi Do Korea Postal Code 14952 Tel 82 31 313 1757 Fax 82 31 314 1989 E mail dmega dmega co kr Home Page http www dmega co kr...

Отзывы: