− Before using the CPM device, the
operator must ascertain that it is in
correct working order and operating
condition. The cables and connectors,
in particular, must be checked for signs
of damage. Damaged parts must be
replaced immediately, before use.
−
Before therapy
, a
test run
consist-
ing of several exercise cycles must
be completed, first without and then
with the patient. Check that all setting
screws are tightened.
− Stop therapy immediately when you
have doubts about the device settings
and/or the therapy protocol.
− Ensure an
anatomically correct setup
of the CPM device
suitable for the
patient to be treated. For this purpose,
check the following settings/positions
(see numbers on device):
1. Height adjustment of motion
element
2. Shoulder width adjustment
of motion element
3. Backrest
4. Length adjustment of upper arm
5. Adjustment of elbow angle
6. Forearm guide
7. Forearm length adjustment
− The CPM device must be stopped be-
fore any of the adjustments 1 through
7 is modified.
− Movement must always be
free of
pain and irritation
.
− The patient must be
fully conscious
during instruction and when using the
device.
− The
choice
of the
therapy param-
eters
to program and of the therapy
protocols to use is
restricted to
the
responsible
physician or therapist
.
The doctor or therapist must decide on
a case-by-case basis whether the de-
vice can be used with each particular
patient.
− The patient must be familiar with
the functions of the ARTROMOT®-S4
programming unit
and the unit must
be
within easy reach
of the patient,
allowing him or her to stop therapy if
needed.
−
Minors
or
persons with reduced
physical, sensory or mental capabili-
ties
must
never be left unattended
during therapy sessions; specialist staff,
a parent or legal guardian must always
be present.
− Write the patient's name on the
pa-
tient chip card
. The card should only
be used for this patient.
If the
patient chip card
is used for
another patient, be sure to delete the
previous patient data
from the card
first (see sections 4.1 and 5.3, para-
graph "New Patient").
Use
original chip cards only
.
− The ARTROMOT®-S4 must only be
used with
accessories
approved by
ORMED.
− Modifications to the medical device
described in this document without
the manufacturer's written consent is
prohibited.
− Do not allow
parts of the body or
objects
(such as blankets, cushions, or
cables) to get caught in the
moving
parts
of the CPM device.