100
5.4.1 Isolated adduction/
abduction
1. Complete the mechanical settings and
adjust the device to the patient as de-
scribed in section 4.2.
2. Then, program the fixed positions for
motors A (anteversion/retroversion) and
C (internal/external rotation) and the
range of motion for motor B (abduction/
adduction) as described in 5.1 in the fol-
lowing order:
• First, program the fixed position for
execution of abduction/adduction in
the anteversion/retroversion plane of
motion (motor A), and then deactivate
the motor.
• Subsequently, program the fixed posi-
tion for execution of abduction/adduc-
tion in the internal/external rotation
plane of motion (motor C), and then
deactivate the motor.
• Finally, program the range of motion
for abduction/adduction (motor B).
3. Save the settings with the
STOP button
,
set the device to the starting position
with the
START button
and initiate
therapy by pressing the
START button
again.
Note!
• For an exclusive adduction/abduction
protocol, motors C (rotation) and A
(anteversion/retroversion) must be
switched off.
• Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
• The stop positions of the different
motors can also be adjusted when
they are deactivated.
5.4.2 Isolated internal/
external rotation
1. Complete the mechanical settings and
adjust the device to the patient as de-
scribed in section 4.2.
2. Then, program the fixed positions for
motors A (anteversion/retroversion) and
B (abduction/adduction) and the range
of motion for motor C (internal/external
rotation) as described in 5.1 in the follow-
ing order:
• First, program the fixed position for
execution of internal/external rotation
in the anteversion/retroversion plane
of motion (motor A), and then deacti-
vate the motor.
• Then, program the fixed position for
execution of internal/external rotation
in the abduction/adduction plane of
motion (motor B), and then deactivate
the motor.
• Finally, program the range of motion
for internal/external rotation (motor
C).
3. Save the settings with the
STOP button
,
set the device to the starting position
with the
START button
, and initiate
therapy by pressing the
START button
again.
Note!
• For an exclusive internal/external
rotation protocol, motors B (abduc-
tion/adduction) and A (anteversion/
retroversion) must be switched off.
• Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
• The fixed stop positions of the differ-
ent motors can also be adjusted when
they are deactivated.
5.4 Application and programming examples