5.6 Examples of use/
sample protocols
5.6.1 Isolated
adduction/abduction
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the
MENU key
on the program‑
ming unit to access level 1 (M1).
3. Press the
parameter key
for
internal
rotation
or external rotation
and
press the
+/- keys
to set the
carriage to the desired rotation position
to be maintained during the isolated
abduction/adduction movements.
4. Next set the desired range of motion
for adduction
and abduction by
first pressing the
parameter key
for
adduction
and set the values with the
+/- keys
.
5. Set the
abduction
value in the same
way
.
Note!
– For an exclusive adduction/abduction
protocol, motor B for the rotation
movement must be turned off as
described in steps 6 to 8 below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the
MENU key
on the program‑
ming unit to access level 3 (M3).
7. Press the Motor B ON/OFF parameter
key
, to activate the parameter.
8. Deactivate motor B by pressing the
parameter key
Motor B ON/OFF
again
or the „–“ (minus) key. The check mark in
the circle next to the symbol must have
disappeared.
9. Save the settings with the
STOP key
, set
the carriage to the starting.
5.6.2 Isolated
internal/external rotation
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the
MENU key
on the program‑
ming unit to access level 1 (M1).
3. Press the
parameter key
for
adduction
or
abduction
and press
the
+/- keys
to set the carriage to the
desired position to be maintained during
the isolated rotation movements.
4. Next set the desired range of motion
for internal/external rotation by first
pressing the
parameter key
for
internal
rotation
and set the values with
the
+/-keys
.
5. Set the
external rotation
value in the
same way
.
Note!
– For an exclusive rotation protocol,
motor A for adduction/ abduction
must be turned off as described in
steps 6 to 8 below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the
MENU key
on the program‑
ming unit to access level 3 (M3).
7. Press the
Motor A ON/OFF
,
parameter key to activate the parameter.
8. Deactivate motor A by pressing the
parameter key
Motor A ON/OFF
again
or the
„–“
(minus) key. The check mark in
the circle next to the symbol must have
disappeared.
9. Save the settings with the
STOP key
,
set the carriage to the starting position
with the
START key
and initiate therapy
by pressing the
START key
again.