Page 19
03/2018
Specification
Units
Ferrite Number FN or Ferrite percentage F%
Range
0 - 115 FN
0 - 80.5% (percent ferrite)
Resolution
0.1 FN or 0.1 F%
Standard: Accuracy (10°C - 30°C)
FN 0 - 10
FN 10 - 30
FN 30 - 100
± 0.5 FN
± 5% of the reading
± 10% of the reading
Air cooled: Accuracy (30°C - 300°C)
FN 0 - 10
FN 10 - 30
FN 30 - 100
± 1 FN
± 10% of the reading
± 20% of the reading
Calibration
Factory calibration to NIST standards
User calibration with transfer standard
Transfer standards
Five transfer standards
Range typically 3 FN to 115 FN
Zero
Automatic on demand
Mode
Average or peak
Storage
1000 records with 4 character record ID
(Characters available 0-9, A-Z and space)
PC Download and Upload
Optional RS232 and USB. Functions include
download of record or part or entire
database and upload of record IDs
Probe P46
Sprung loaded sleeve
Pencil style, 15 mm dia. length 130 mm
Sensing area 4 mm x 6 mm
Probe P10
Pencil style, 15mm diameter length 120mm
Sensing area 10mm diameter
Operating temperature
Standard 0°C - 40°C
Air cooled 0°C - 300°C
Air cooled operation (P10ac)
Air line supply regulated to 1 bar maximum
Oil filtered supply desirable
Air temperature: ambient (10°C - 30°C)
Humidity
Non-condensing
Weight in case
1.25 kg
Case dimensions
210 x 100 x 45 mm
Support
Call/email Diverse for support for use of the
instrument or assistance with unusual
application.