Software -
The instrument provides for storage of up to 1000 sample records.
Each record is accompanied by a 4 character ID which is used to
identify the place where the measurement is taken. If you do not
want to use the ID then on sampling the sample number is used.
The serial interface is upgraded to 19200 Baud allowing for quicker
downloads of the full database.
Records are stored as a fixed field format: xxxyyyyyFNzzzz
where xxx is the record number 000 - 999
yyyyyFN is the ferrite number 000.0 - 115.0
zzzz is the ID for user z is any alphanumeric character
From the computer the following additional functions are available:
Key
Function
Pnnn
Moves the record pointer to nnn: 000 - 999
C
Clear the database (only available from other
terminals, use drop down menu for this function
with serial software)
R
Download record at current pointer and increment
pointer
S
Get status from the instrument - used during
transfers to ensure new commands are not sent
before the last one finishes.
Software -