13.0 Information on electro-
magnetic immunity
Guidelines and manufacturer's declaration –
electromagnetic immunity
The BSM 382 is intended for use in the electromagnetic
environment specified below. The customer or the user
of the BSM 382 should assure that it is used in such an
environment.
Immunity
test
IEC
60601-
test
level
Confor
mity
level
Electromagnetic
environment –
guidelines
Electrostati
c discharge
(ESD)
according
to IEC
61000-4-2
± 6 kV
contact
discharge
± 8 kV
air
discharge
Not
applicabl
e
± 8 kV
air
discharg
e
Floors should be
wood, concrete or
laid with ceramic
tiles. If floors are
covered with
synthetic material,
the relative humidity
should be at least 30
%.
Electrical
fast
transient/b
urst
according
to IEC
61000-4-4
± 2 kV
for mains
lines
± 1 kV
for
input/out
put lines
Not
applicabl
e
Mains power quality
should be that of a
typical residential,
commercial or
hospital
environment.
39
ELECTRICAL IMMUNITY
GB