Settings
Sounds Setup
Sp02 Alm Hi 100
Sp02 Alm Lo 90
PR Alm Hi 130
PR Alm Lo 50
+/- +
Exit
*
Fig. 11
29
move the cursor down to the PR Alm Hi entry to
adjust the upper alarm limit next to the star-
shaped cursor. To do so, hold down the M
configuration button until the lowest limit of 5 is
displayed. As you hold down the button , the
displayed limit increases in steps of five to a
maximum value of 250 before starting at 5 again.
Release the button when the desired value
(e.g. 130, refer to fig. 11) is reached and move
the star-shaped cursor down to the PR Alm Lo
entry by pressing the configuration button
again. Adjust the lower alarm limit next to the
star-shaped cursor in the same way as the upper
alarm limit. Make sure that the upper limit is
higher than the lower limit. To save the selected
settings, move the star-shaped cursor to the Exit
entry by pressing the configuration button and
hold the button down until the measurements
are displayed again. The settings are now saved.