43
Each output card is supplied with an adhesive label that indicates the wiring connections of each option. To help
identifying each terminal, this label should be placed in the lower side of the meter case, beside the basic functions
label.
NOTE
: In case that the outputs are used to drive inductive loads, it is recommended to add an RC network between
the coil terminals (preferably) or between the relay contacts to limit electromagnetic effects.
9.1.4 – technical specifications
CHARACTERISTICS
2RE OPTION
4RE OPTION
MAX.CURRENT(RESISTIVE LOAD) ...................................................... 8 A ......................................... 5 A
MAX.POWER ................................................................. 2000 VA / 192 W .................... 1250 VA / 150 W
MAX.VOLTAGE ........................................................... 250 VAC / 150 VDC ................. 277 VAC / 125 VDC
CONTACT RESISTANCE ............................................................ Max. 3m
Ω
............................. Max. 30m
Ω
SWITCHING TIME .................................................................. Max. 10ms .............................. Máx. 10ms
4OP and 4OPP OPTION
MAX VOLTAGE ............................................................................. 50 VDC
MAX CURRENT ............................................................................ 50 mA
LEAKAGE CURRENT ........................................................... 100 µA (max.)
SWITCHING TIME ................................................................ 1 ms (max.)
43
Each output card is supplied with an adhesive label that indicates the wiring connections of each option. To help
identifying each terminal, this label should be placed in the lower side of the meter case, beside the basic functions
label.
NOTE
: In case that the outputs are used to drive inductive loads, it is recommended to add an RC network between
the coil terminals (preferably) or between the relay contacts to limit electromagnetic effects.
9.1.4 – technical specifications
CHARACTERISTICS
2RE OPTION
4RE OPTION
MAX.CURRENT(RESISTIVE LOAD) ....................................................... 8 A ........................................ 5 A
MAX.POWER .................................................................. 2000 VA / 192 W .................... 1250 VA / 150 W
MAX.VOLTAGE ........................................................... 250 VAC / 150 VDC ................. 277 VAC / 125 VDC
CONTACT RESISTANCE ............................................................ Max. 3m
Ω
............................. Max. 30m
Ω
SWITCHING TIME ................................................................... Max. 10ms ............................. Máx. 10ms
4OP and 4OPP OPTION
MAX VOLTAGE ............................................................................. 50 VDC
MAX CURRENT ............................................................................ 50 mA
LEAKAGE CURRENT .......................................................... 100 µA (max.)
SWITCHING TIME ................................................................. 1 ms (max.)