24
CALn
(Calibrate Scale)
Step 1.
The display will change to show CALn.
Step 2.
To skip calibration and save the previous settings, press the “ ” key to exit calibration. The
scale will turn off automatically. Proceed to Step 14.
Step 3.
Otherwise, press the or keys to select
y
(Y = start calibration). Note that you can still
select
n
(n = exit calibration), skip calibration and turn the scale off.
Step 4.
Press the “ ” key.
Step 5.
The display will change to show the internal count.
Step 6.
Make certain the scale platform is empty and free of debris.
Step 7.
Press the “ ” key to save the zero reference point.
Step 8.
The display will change to show the first calibration point
100 kg
or
200 lb
Step 9.
Place the first calibrated test weight on the scale platform.
Step 10.
When reading is stable, the first weight reading will be saved and the display will change to
show the second calibration point
200 kg
or
400 lb.
Step 11.
Place the next calibrated test weight on the scale platform.
Step 12.
When reading is stable, the second weight reading will be saved.
Step 13.
The display will change to show
PASS
and then automatically shut the scale off.
Step 14.
The scale is now ready for operation.
Calibration
(cont.)