background image

TECHNICAL MANUAL

9

GATERD.SE

5. Replacement of remote control

Incorrect remote control is easily changed by releasing the contact / remote control on the left pillar - new 

remote control is mounted in the same way.

NOTE: It is important that contact is locked with the clamp lock to ensure IP X5.

6. Calibration of actuators

The calibration of the actuators (height / lowering only) is normally done fully automatically each time the chair 
reaches its end positions (min height, max tilt).

If the chair still does not go in parallel or if the tilt function does not seem to work, an ACTIVE calibration can be 
done. This is done by holding down <button-slow / up> + <button-slow / down> simultaneously.

After about 15 seconds the chair will be moved to its end positions ie. my height and maximum slope 
backwards. Hold down the buttons until the end positions are reached. Release the buttons and wait 5 
seconds. The chair is now calibrated and can be used again.

 

Calibration should NOT be done with user in chair. Note that all 3 actuators move during the calibration  

 

phase. Otherwise, cancel the calibration immediately and contact Customer Service / GATE (or your  

 

local dealer).

7. Reset service indicator

The remote control has a service indicator for periodic maintenance reminder (min once a year). The service 
indicator measures voltage and is activated after 11 months (flashes every 4 seconds). After 12 months the 

indicator will glow steadily.

The indicator is reset by holding down buttons on the remote control (top line, <fast up>, <fast down>) for 30 

seconds.

Содержание Gate EASY SoftBack

Страница 1: ...A DIRECT HEALTHCARE GROUP COMPANY G ATERD S E EASY SoftBack Technical Manual...

Страница 2: ...EASY SOFTBACK GATERD SE 2...

Страница 3: ...Unequalled aid for personnel while also providing great comfort for the user for example its ability to tilt really helps take the strain EASY is available with our new soft Comfort seat with an open...

Страница 4: ...left right 11 58 296 1 Footrest 12 58 281 1 Bottom frame 13 57 049 RF Wheel with bolt fully braked 13 57 040 RF Wheel with bolt directional lock and fully braked 14 58 360 Cover 15 57 793 End plug 16...

Страница 5: ...TECHNICAL MANUAL 5 GATERD SE...

Страница 6: ...216 Infill 56 217 Waist belt Thigh belt Hygiene chairs 56 218 Slides 56 220 EASY Heel strap kit pair Easy 56 221 Width kit 50mm Easy SB 56 227 L Calf rest left 56 227 R Calf rest right 56 229 Chest be...

Страница 7: ...with a plastic bag or the like that can withstand the wa ter pressure of the washing machine Place remote control control box or connector female so that running water cannot penetrate Seat can be rem...

Страница 8: ...are whole these must be replaced if they are damaged 3 The seat soft seat is complete and without cracks No sharp edges may be present 4 Footrest plates are whole and possible to adjust in the intend...

Страница 9: ...simultaneously After about 15 seconds the chair will be moved to its end positions ie my height and maximum slope backwards Hold down the buttons until the end positions are reached Release the butto...

Страница 10: ...Off button see below Wait a minute To activate the battery press its On button see below Make sure the battery indicator light is on or flashing 1a Indicator shows green Calibrate actuator see page 7...

Страница 11: ...rranty We normally provide a 2 year warranty on manufacturing defects does not apply to wear details or battery The battery is given a 6 month warranty on manufacturing defects For more information or...

Страница 12: ...Lift 100kg pelare 150 kg Stroke 400 mm 75 mm Installation dimensions 377 mm 198 mm Protection class N A IP66W Protection class pillars IPX5 Speed without load 32 mm s 11 5 mm s Speed full load 19 mm s...

Страница 13: ...y removing the protective cap 4 screws Control box with battery with removed cover Actuators 2x up down and 1x tilt The actuators are connected directly to the control box as below Remote control and...

Страница 14: ...Hygiene Chair s electrical components or cabling with original parts from supplier Failure to do so may affect the product s EMC approval Avoid placing using the chair less than 0 35m from other medi...

Страница 15: ...differential mode n a for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input l...

Страница 16: ...than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies...

Страница 17: ...TECHNICAL MANUAL 17 GATERD SE 15 Dimensions All dimensions in millimeters and with 125 mm wheels When using Double wheels 100mm width over wheels increases by 30mm...

Страница 18: ...EASY SOFTBACK GATERD SE 18...

Страница 19: ...TECHNICAL MANUAL 19 GATERD SE...