TECHNICAL MANUAL

15

GATERD.SE

The product is intended for use in the electromagnetic environment specified below. The customer

or the user of the product should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic 
environment - guidance

Electrostatic discharge

(ESD)
IEC 61000-4-2

+/- 8 kV contact
+/- 2,4,8,15 kV air

+/- 8 kV contact
+/- 2,4,8,15 kV air

Floors should be wood,

concrete or ceramic
tile.

If floors are covered
with synthetic material,

the relative humidity

should be at least 30

%.

Electrical fast transient /
Burst

IEC 61000-4-4

+/- 2 kV for power

supply lines

+/- 1 kV for input/

output lines

+/- 2 kV for power

supply lines

n/a. for input/output
lines

Mains power quality
should be that of a

typical commercial or
hospital environment.

Surge

IEC 61000-4-5

+/- 1 kV differential

mode

+/- 2 kV common

mode

+/- 1 kV differential

mode

n/a. for common mode

Mains power quality
should be that of a

typical commercial or
hospital environment.

Voltage dips, short

interruptions and voltage
variations on power 
supply
input lines

IEC 61000-4-11

<5 % UT
(>95 % dip in UT)

 

for 0,5 cycle

40 % UT
(60 % dip in UT)

for 5 cycles

70 % UT
(30 % dip in UT)

for 25 cycles

<5 % UT
(>95 % dip in UT))

for 5 sec

<5 % UT
(>95 % dip in UT)
for 0,5 cycle

40 % UT
(60 % dip in UT)

for 5 cycles

70 % UT
(30 % dip in UT)

for 25 cycles

<5 % UT
(>95 % dip in UT))

for 5 sec

Mains power quality
should be that of

a typical commercial
or hospital environment.

If the user of the
[Equipment or System]
requires continued

operation during power

mains interruptions, it is

recommended that the

[Equipment or System]
be powered from an
uninterruptible power
supply or battery.

Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8

30 A/m

30 A/m

Power frequency 
magnetic fields should be

at levels characteristic
of a typical location in
a typical commercial or
hospital environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.

Содержание Gate EASY SoftBack

Страница 1: ...A DIRECT HEALTHCARE GROUP COMPANY G ATERD S E EASY SoftBack Technical Manual...

Страница 2: ...EASY SOFTBACK GATERD SE 2...

Страница 3: ...Unequalled aid for personnel while also providing great comfort for the user for example its ability to tilt really helps take the strain EASY is available with our new soft Comfort seat with an open...

Страница 4: ...left right 11 58 296 1 Footrest 12 58 281 1 Bottom frame 13 57 049 RF Wheel with bolt fully braked 13 57 040 RF Wheel with bolt directional lock and fully braked 14 58 360 Cover 15 57 793 End plug 16...

Страница 5: ...TECHNICAL MANUAL 5 GATERD SE...

Страница 6: ...216 Infill 56 217 Waist belt Thigh belt Hygiene chairs 56 218 Slides 56 220 EASY Heel strap kit pair Easy 56 221 Width kit 50mm Easy SB 56 227 L Calf rest left 56 227 R Calf rest right 56 229 Chest be...

Страница 7: ...with a plastic bag or the like that can withstand the wa ter pressure of the washing machine Place remote control control box or connector female so that running water cannot penetrate Seat can be rem...

Страница 8: ...are whole these must be replaced if they are damaged 3 The seat soft seat is complete and without cracks No sharp edges may be present 4 Footrest plates are whole and possible to adjust in the intend...

Страница 9: ...simultaneously After about 15 seconds the chair will be moved to its end positions ie my height and maximum slope backwards Hold down the buttons until the end positions are reached Release the butto...

Страница 10: ...Off button see below Wait a minute To activate the battery press its On button see below Make sure the battery indicator light is on or flashing 1a Indicator shows green Calibrate actuator see page 7...

Страница 11: ...rranty We normally provide a 2 year warranty on manufacturing defects does not apply to wear details or battery The battery is given a 6 month warranty on manufacturing defects For more information or...

Страница 12: ...Lift 100kg pelare 150 kg Stroke 400 mm 75 mm Installation dimensions 377 mm 198 mm Protection class N A IP66W Protection class pillars IPX5 Speed without load 32 mm s 11 5 mm s Speed full load 19 mm s...

Страница 13: ...y removing the protective cap 4 screws Control box with battery with removed cover Actuators 2x up down and 1x tilt The actuators are connected directly to the control box as below Remote control and...

Страница 14: ...Hygiene Chair s electrical components or cabling with original parts from supplier Failure to do so may affect the product s EMC approval Avoid placing using the chair less than 0 35m from other medi...

Страница 15: ...differential mode n a for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input l...

Страница 16: ...than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies...

Страница 17: ...TECHNICAL MANUAL 17 GATERD SE 15 Dimensions All dimensions in millimeters and with 125 mm wheels When using Double wheels 100mm width over wheels increases by 30mm...

Страница 18: ...EASY SOFTBACK GATERD SE 18...

Страница 19: ...TECHNICAL MANUAL 19 GATERD SE...

Страница 20: ...GATER D SE Gate Rehab Development AB Industrigatan 2 Reftele Sverige S 330 21 T 0371 318 00 F 0371 318 01 E info gaterd se LIT 00116 Issue 1 Date January 2020 A DIRECT HEALTHCARE GROUP COMPANY...

Отзывы: