The product is intended for use in the electromagnetic environment specified below.

The customer or the user of the product should assure that it is used in such an environment.

Immunity test

IEC 60601 test 

level

Compliance level

Electromagnetic environment - guidance

Portable and mobile RF communications
equipment should be used no closer than
0,35m to any part of the product, 

including

cables, than the recommended separation
distance calculated from the equation 
applicable
to the frequency of the transmitter.

Recommended separation distance

d = 1,2√p        80 MHz to 800 MHz

d = 2,3√p        800 MHz to 2.5 GHz

where P is the maximum output power
rating of the transmitter in watts (W) 

according to the transmitter manufacturer 
and d is the recommended separation 

distance in meters (m).

Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the 
compliance level in each frequency range. 
b

Interference may occur in the vicinity of
equipment marked with the following
symbol.

NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflected from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which the 
product is used exceeds the applicable RF compliance level above, the product should be observed to verify 
normal operation. If abnormal performance is observed, additional measures may be necessary, such as 

reorienting or relocating the product.

b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.

Содержание Gate EASY SoftBack

Страница 1: ...A DIRECT HEALTHCARE GROUP COMPANY G ATERD S E EASY SoftBack Technical Manual...

Страница 2: ...EASY SOFTBACK GATERD SE 2...

Страница 3: ...Unequalled aid for personnel while also providing great comfort for the user for example its ability to tilt really helps take the strain EASY is available with our new soft Comfort seat with an open...

Страница 4: ...left right 11 58 296 1 Footrest 12 58 281 1 Bottom frame 13 57 049 RF Wheel with bolt fully braked 13 57 040 RF Wheel with bolt directional lock and fully braked 14 58 360 Cover 15 57 793 End plug 16...

Страница 5: ...TECHNICAL MANUAL 5 GATERD SE...

Страница 6: ...216 Infill 56 217 Waist belt Thigh belt Hygiene chairs 56 218 Slides 56 220 EASY Heel strap kit pair Easy 56 221 Width kit 50mm Easy SB 56 227 L Calf rest left 56 227 R Calf rest right 56 229 Chest be...

Страница 7: ...with a plastic bag or the like that can withstand the wa ter pressure of the washing machine Place remote control control box or connector female so that running water cannot penetrate Seat can be rem...

Страница 8: ...are whole these must be replaced if they are damaged 3 The seat soft seat is complete and without cracks No sharp edges may be present 4 Footrest plates are whole and possible to adjust in the intend...

Страница 9: ...simultaneously After about 15 seconds the chair will be moved to its end positions ie my height and maximum slope backwards Hold down the buttons until the end positions are reached Release the butto...

Страница 10: ...Off button see below Wait a minute To activate the battery press its On button see below Make sure the battery indicator light is on or flashing 1a Indicator shows green Calibrate actuator see page 7...

Страница 11: ...rranty We normally provide a 2 year warranty on manufacturing defects does not apply to wear details or battery The battery is given a 6 month warranty on manufacturing defects For more information or...

Страница 12: ...Lift 100kg pelare 150 kg Stroke 400 mm 75 mm Installation dimensions 377 mm 198 mm Protection class N A IP66W Protection class pillars IPX5 Speed without load 32 mm s 11 5 mm s Speed full load 19 mm s...

Страница 13: ...y removing the protective cap 4 screws Control box with battery with removed cover Actuators 2x up down and 1x tilt The actuators are connected directly to the control box as below Remote control and...

Страница 14: ...Hygiene Chair s electrical components or cabling with original parts from supplier Failure to do so may affect the product s EMC approval Avoid placing using the chair less than 0 35m from other medi...

Страница 15: ...differential mode n a for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input l...

Страница 16: ...than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies...

Страница 17: ...TECHNICAL MANUAL 17 GATERD SE 15 Dimensions All dimensions in millimeters and with 125 mm wheels When using Double wheels 100mm width over wheels increases by 30mm...

Страница 18: ...EASY SOFTBACK GATERD SE 18...

Страница 19: ...TECHNICAL MANUAL 19 GATERD SE...

Страница 20: ...GATER D SE Gate Rehab Development AB Industrigatan 2 Reftele Sverige S 330 21 T 0371 318 00 F 0371 318 01 E info gaterd se LIT 00116 Issue 1 Date January 2020 A DIRECT HEALTHCARE GROUP COMPANY...

Отзывы: