dipper KN-201 Скачать руководство пользователя страница 8

Press and hold 

[Bean/Power]

 button for 3 seconds in the power-on state, 

"PowerOff" will show on the screen. The equipment will enter into shutdown 

state. 

 

 

Coffee sample placement 

Since coffee samples that have just been roasted or are still temperature-

charged will release infrared light. This may interfere with the sensor resulting 

in inaccurate measurement. Make sure the coffee sample has cooled down to 

room temperature before using. 

When the sample coffee is not properly placed, the measurement result 

can become inaccurate. The placement density of the sample and whether or 

not it levels with the edge of the cup when filling and flattening the sample 

cup  are  the  two  most  important  points  that  will  affect  the  measurement 

accuracy. When placing coffee in the sample cup, fill as much as possible to 

reduce any gap in the content. Then use a scraping tool to flatten the top to 

the height of the cup edge level without special weighing. 

   

 

 

 

The degree of coffee grinding will also affect the measurement result. The 

SCA cupping standards requires 70 to 75% of ground must pass through 20 

mesh sieve. The ground is then flattened to fill the sample cup. Make sure 

there is no gap in the content and is level with the cup edge. If there is no 

correct sieve available, you can fix each grinding scale for comparison. This will 

Coffee bean

 

Coffee ground

 

Содержание KN-201

Страница 1: ...Coffee Roast Analyzer KN 201 User manual Made in Taiwan Version 1 1 6...

Страница 2: ...technology complies with the SCA Roast Color Classification System which enables a coffee roaster accurately measuring the coffee beans Agtron number and the roasting degree name This product uses th...

Страница 3: ...low degree However as the market evolved this type of measuring method is gradually replaced by other measuring equipment that turned to the primary use of assessing the degree of roasting Agtron Gour...

Страница 4: ...s otherwise it will cause damage This product is a precision optical instrument If the sensor or light source is contaminated the accuracy of the measurement will be affected Keep the use and storage...

Страница 5: ...und measurements Self adaptability design error automatic compensation Comply with standard SCA Agtron scales Gourmet and Commercial Display Roast degree name SCA and Common support 200 data logs of h...

Страница 6: ...e short press to start measurement for ground coffee Micro USB To connect the USB charger Charging indicator Orange light indicates charging is in progress and green light indicates charging is comple...

Страница 7: ...tes that charging is in progress When showing Full Charged means the battery is fully charged and the charging will automatically stop This product supports charging and operating the measurement work...

Страница 8: ...er or not it levels with the edge of the cup when filling and flattening the sample cup are the two most important points that will affect the measurement accuracy When placing coffee in the sample cu...

Страница 9: ...er and the corresponding two roasting degree names The name of the SCA definition shows above the Agtron number the common name shows under Over Range is shown when the measured value is outside the r...

Страница 10: ...side of the calibration plate on a flat surface then place the KN 201 on the calibration plate Make sure there is a solid fit all around Press and hold Ground Setting button for more than 3 seconds to...

Страница 11: ...se accuracy When carrying out calibration the ambient temperature should be kept in the range of 25 5 C Temperature outside of this range may affect the accuracy of the calibration After the calibrati...

Страница 12: ...g to select the item options The options are displayed in the table below Item Option Description 1 Auto Power Off 1 Always ON 2 5 mins 3 10 mins 4 30 mins 5 60 mins The default is 30 minutes When the...

Страница 13: ...e the charging cable before cleaning Exterior cleaning The body shell and sample cup plate can be wiped with a micro wet fine fiber cloth and then with a dry cloth to remove the surface moisture Pleas...

Страница 14: ...it use a small amount of soft dishwashing liquid and then rinse with water dry quickly with soft paper towels or cloth The process must be gentle to avoid surface scratches Do not clean light sources...

Страница 15: ...ED Sensor Multi point phototransistors Warm up time 30 seconds Measure range 0 150 Agtron Gourmet number Accuracy 99 Data records 200 records Battery High capacity rechargeable polymer lithium battery...

Отзывы: