The  Specialty  Coffee  Association  (SCA),  a  body  merged  from  European 

Specialty Coffee Association (SCAE)  and American Specialty Coffee Association 

(SCAA), promotes using

 

infrared analysis to assess the degree of coffee roast. The 

method  used  is  called  an

 

Agtron  number.  A  lower  Agtron  number

 

indicates  a 

higher caramelization and a deeper roasting degree. On the other hand, a higher 

number would indicate a lower caramelization and a lighter roast degree. This 

quantitative  way  can  assist  roasters  in  determining  the  degree  of  coffee  roast 

more  precisely  which  is  a  great  aid  for  them  in  managing  the  quality  and 

consistency of their products. 

This product is a professional precision instrument for measuring the degree 

of coffee roast. It is suitable for both coffee beans and ground coffee. This product 

provides a user friendly interface with an excellent quality exterior. The built in 

optical technology complies with the SCA Roast Color Classification System which 

enables a coffee roaster accurately measuring the coffee beans Agtron number 

and the roasting degree name. 

 

 

This product uses the same principle of near-infrared sensing as Agtron. When 

the infrared wavelength is between 780 to 1500nm, it is called a near-infrared due 

to its proximity to visible wavelengths. Many researches point out that the deeper 

the coffee roasting degree, the higher the carbonization degree of soluble organic 

matter  in  coffee  beans,  and  the  higher  the  absorption  of  near-infrared  light. 

Therefore, when the measurement instrument emits near-infrared to the coffee 

to  be  tested,  then  measure  the   reflected  energy.  The  lower  the  reflection 

indicates  the  deeper  the  roasting  degree;  the  higher  the  reflective  amount 

indicates  a  shallower  roasting  degree.  SCA  converts  this  reflection  into  a 

numerical value and calls it an Agtron number. 0 means that the soluble organic 

matter in coffee beans has been 100% carbonized, and 105 represents only 3-5% 

coffee sucrose. 

Product description

 

Analysis principle

 

Содержание KN-201

Страница 1: ...Coffee Roast Analyzer KN 201 User manual Made in Taiwan Version 1 1 6...

Страница 2: ...technology complies with the SCA Roast Color Classification System which enables a coffee roaster accurately measuring the coffee beans Agtron number and the roasting degree name This product uses th...

Страница 3: ...low degree However as the market evolved this type of measuring method is gradually replaced by other measuring equipment that turned to the primary use of assessing the degree of roasting Agtron Gour...

Страница 4: ...s otherwise it will cause damage This product is a precision optical instrument If the sensor or light source is contaminated the accuracy of the measurement will be affected Keep the use and storage...

Страница 5: ...und measurements Self adaptability design error automatic compensation Comply with standard SCA Agtron scales Gourmet and Commercial Display Roast degree name SCA and Common support 200 data logs of h...

Страница 6: ...e short press to start measurement for ground coffee Micro USB To connect the USB charger Charging indicator Orange light indicates charging is in progress and green light indicates charging is comple...

Страница 7: ...tes that charging is in progress When showing Full Charged means the battery is fully charged and the charging will automatically stop This product supports charging and operating the measurement work...

Страница 8: ...er or not it levels with the edge of the cup when filling and flattening the sample cup are the two most important points that will affect the measurement accuracy When placing coffee in the sample cu...

Страница 9: ...er and the corresponding two roasting degree names The name of the SCA definition shows above the Agtron number the common name shows under Over Range is shown when the measured value is outside the r...

Страница 10: ...side of the calibration plate on a flat surface then place the KN 201 on the calibration plate Make sure there is a solid fit all around Press and hold Ground Setting button for more than 3 seconds to...

Страница 11: ...se accuracy When carrying out calibration the ambient temperature should be kept in the range of 25 5 C Temperature outside of this range may affect the accuracy of the calibration After the calibrati...

Страница 12: ...g to select the item options The options are displayed in the table below Item Option Description 1 Auto Power Off 1 Always ON 2 5 mins 3 10 mins 4 30 mins 5 60 mins The default is 30 minutes When the...

Страница 13: ...e the charging cable before cleaning Exterior cleaning The body shell and sample cup plate can be wiped with a micro wet fine fiber cloth and then with a dry cloth to remove the surface moisture Pleas...

Страница 14: ...it use a small amount of soft dishwashing liquid and then rinse with water dry quickly with soft paper towels or cloth The process must be gentle to avoid surface scratches Do not clean light sources...

Страница 15: ...ED Sensor Multi point phototransistors Warm up time 30 seconds Measure range 0 150 Agtron Gourmet number Accuracy 99 Data records 200 records Battery High capacity rechargeable polymer lithium battery...

Отзывы: