Charging 

The new machine is not fully charged after it left the factory. A full charge 

is recommended before first use. Connect one end of the USB charging cable 

to the charger, and the other end of the micro USB to the USB charging port 

of KN-201, then plug the charger into an AC110/220V socket. When orange 

light  shows  up,  indicates  charging  is  in  progress;  and  green  light  indicates 

charging is complete. It takes about 3 hours to get fully charged when the 

battery  is  completely  discharged.  It  can  be  powered  on  continuously  for  8 

hours after a fully charge. 

When  charging  in  the  power-on  state,  a  plug  symbol  with  the  word  C 

appears on the upper right hand corner of the display screen. The C word will 

be  replaced  with  a  simple  plug  symbol  when  fully  charged.  When  the 

equipment is turned off, tap on any key to wake up the display. If the screen 

displays  "Battery  Charging",  indicates  that  charging  is  in  progress.  When 

showing "Full Charged", means the battery is fully charged and the charging 

will  automatically  stop.  This  product  supports  charging  and  operating  the 

measurement work at the same time. 

 

  When  the  battery  is  completely  discharged,  "battery  Empty"  is 

displayed  and  the  equipment  will  shut  down  automatically.  It  is 

recommended to plug in the charger to charge the battery before using the 

equipment to avoid damaging the battery. 

 If  the  product  is  not  frequently  used,  we  recommend  charging  the 

battery at least once a month, to avoid damaging the lithium battery due to 

low battery voltage. 

 

 

Power switch 

Press  and  hold 

[Bean/Power]

  button  for  1  second,  the  screen  will  first 

show the product logo. Then it will enter into the 30 seconds warm-up screen. 

To ensure measurement accuracy, it is recommended to wait for the warm-up 

process to complete before starting a measurement operation. If skipping the 

warm-up  process  is  really  necessary,  press  the  two  keys  on  the  panel 

[Bean/Power]

[Ground/Setting]

  simultaneously,  "Quit"  will  show  on  the 

screen. 

Operation method

 

Содержание KN-201

Страница 1: ...Coffee Roast Analyzer KN 201 User manual Made in Taiwan Version 1 1 6...

Страница 2: ...technology complies with the SCA Roast Color Classification System which enables a coffee roaster accurately measuring the coffee beans Agtron number and the roasting degree name This product uses th...

Страница 3: ...low degree However as the market evolved this type of measuring method is gradually replaced by other measuring equipment that turned to the primary use of assessing the degree of roasting Agtron Gour...

Страница 4: ...s otherwise it will cause damage This product is a precision optical instrument If the sensor or light source is contaminated the accuracy of the measurement will be affected Keep the use and storage...

Страница 5: ...und measurements Self adaptability design error automatic compensation Comply with standard SCA Agtron scales Gourmet and Commercial Display Roast degree name SCA and Common support 200 data logs of h...

Страница 6: ...e short press to start measurement for ground coffee Micro USB To connect the USB charger Charging indicator Orange light indicates charging is in progress and green light indicates charging is comple...

Страница 7: ...tes that charging is in progress When showing Full Charged means the battery is fully charged and the charging will automatically stop This product supports charging and operating the measurement work...

Страница 8: ...er or not it levels with the edge of the cup when filling and flattening the sample cup are the two most important points that will affect the measurement accuracy When placing coffee in the sample cu...

Страница 9: ...er and the corresponding two roasting degree names The name of the SCA definition shows above the Agtron number the common name shows under Over Range is shown when the measured value is outside the r...

Страница 10: ...side of the calibration plate on a flat surface then place the KN 201 on the calibration plate Make sure there is a solid fit all around Press and hold Ground Setting button for more than 3 seconds to...

Страница 11: ...se accuracy When carrying out calibration the ambient temperature should be kept in the range of 25 5 C Temperature outside of this range may affect the accuracy of the calibration After the calibrati...

Страница 12: ...g to select the item options The options are displayed in the table below Item Option Description 1 Auto Power Off 1 Always ON 2 5 mins 3 10 mins 4 30 mins 5 60 mins The default is 30 minutes When the...

Страница 13: ...e the charging cable before cleaning Exterior cleaning The body shell and sample cup plate can be wiped with a micro wet fine fiber cloth and then with a dry cloth to remove the surface moisture Pleas...

Страница 14: ...it use a small amount of soft dishwashing liquid and then rinse with water dry quickly with soft paper towels or cloth The process must be gentle to avoid surface scratches Do not clean light sources...

Страница 15: ...ED Sensor Multi point phototransistors Warm up time 30 seconds Measure range 0 150 Agtron Gourmet number Accuracy 99 Data records 200 records Battery High capacity rechargeable polymer lithium battery...